Section 4 Corrections
VM-21P 3-Wire LVDT Signal Conditioner Instruction Manual
- 28 -
ATTENTION
Section 4 Corrections
To maintain the performance of this product, make corrections in the following instances.
●
When turning on the power supply for the first time
●
When 1 year has passed since the previous correction
●
When the performance of the product has dropped for some reason
4.1 Correction
preparations
●
The T.P. output signal is not insulated from the input signal.
●
The common for the T.P. output signal should be insulated from the ground to prevent
noise.
Place a T.P.output signal receiver as shown below, and make corrections.
Refer to “4.2 Correcting the product” (
☞
Page 29) for the procedure for making corrections.
When using a signal conditioner socket
When using a backplane unit
Point
The example in the illustration shows a connection to Slot 1. If a
connection is made to a different slot, make connections in the same way
to terminals 5 and 6.
Digital multimeter
Digital multimeter