3. Operation 

1. Determine the head end of cable to be identified, and pick out the one (or two, depending 

on the testing mode adopted) from the target optical cables to connect the testing equipment; 

 

2. Connect the optical fiber to open testing equipment at the head end of the target cable. It 

shall be noted that the optical interface is of FC/APC for the testing equipment, and please 

use the transfer jumper for connection if the optical connector is not of FC/APC for the 

target cable termination. Our standard configuration includes the jumper from FC/APC to 

FC/UPC; 

 

3. The optical fiber connector shall be of PC at the end of target cable in order to enhance 

the strength of reflection signal, and reflector may be connected at the end if necessary; 

 

4. The light power value received shall be greater than -55dBm at the testing equipment, 

otherwise please check whether the connection is correct or whether the line loss is too 

large; 

 

5. Unfasten at least 1m of the original binding fiber bundle at the identified cable, and 

number each cable with tape; 

 

6. Use the headset of the testing equipment at the head end of  target cable to listen to the 

voice by tapping the cable, and adjust the sound volume to the appropriate value through the 

adjusting knob on the headset; 

 

7. Use the metal rod of screwdriver to tap each cable with the same strength at the cable to 

be identified, and use the headset of the testing equipment at the head end of target cable to 

listen to and record the voices of tapping different cables. It shall be noted at this step that: 

in order to prevent the vibration from transmitting to other cables when tapping one, a 

sponge or soft cloth shall be used at the binding to separate the cable tapped with other 

cables. The same effect may be achieved if one person holds the cable with his hands and 

another taps the cable; 

 

8. The above step shall be repeated by reducing the tapping strength if the voices by tapping 

different cables are almost the same. In most cases, the cable shall only be tapped slightly 

with finger; 

 

9. Analyze which one is the target cable. There shall be an optical cable of which the voice 

Summary of Contents for OFI-60

Page 1: ...OFI 60 Optical Cable Identifier User s Manual Shineway Technologies Inc All rights reserved...

Page 2: ...FI 60 Optical Cable Identifier is a laser device user should always avoid direct eye exposure to the laser output Using microscope or magnifier to observe the laser output should also be avoided laser...

Page 3: ...Always avoid looking directly into the optical output port when the instrument is working Always replace protective dust cap on the detector port when the instrument is not in use Always avoid lookin...

Page 4: ...ERATION 4 2 1 FOREWORD 4 2 2 1 INSTRUMENT DETAILS 4 2 2 2 Keypad Instruction to Instrument Panel and Functions 5 2 3 USE OF BATTERY 7 2 4 CONNECTOR CLEANING 7 3 OPERATION 8 4 MAINTENANCE AND CALIBRATI...

Page 5: ...st Upon receiving the instrument please check for any obvious signs of physical damage that may have occurred during shipment Report any damage to the shipping agent or the representative of Shineway...

Page 6: ...urately identify the target cable OFI 60 has a high sensitivity strong resistance to electromagnetic interference easy operation and large dynamic range it plays a very important role for computer roo...

Page 7: ...and for Internet access 2 USB and mini USB interfaces used for equipment calibration communication and software upgrades etc 3 Headset interface used to monitor the audio signal produced by the vibrat...

Page 8: ...e OCID switch between running and stop states After pressing this button the instrument is in the testing mode and the status bar of instrument display shows running with real time waveform while the...

Page 9: ...esents the run stop button making the OCID switch between running and stop states After pressing this button the instrument is in the testing mode and the status bar of instrument display shows runnin...

Page 10: ...hen LCD display becomes dim laser source output will become unstable and power meter measurement will be inaccurate Please charge as soon as possible when battery is low to ensure accurate measurement...

Page 11: ...ten at least 1m of the original binding fiber bundle at the identified cable and number each cable with tape 6 Use the headset of the testing equipment at the head end of target cable to listen to the...

Page 12: ...ding to the record results 4 Maintenance and Calibration 4 1 Cleaning of the connectors Keep the cleanness of connectors and dust caps The detector needs to be cleaned timely 4 2 Calibration Requireme...

Page 13: ...ce or accident The Provider reserves the right to make changes to any of its products at any time without having to replace or change previously purchased units 5 3 Warranty Registration A warranty re...

Page 14: ...5 Contact Customer Service Please check our web site www shinewaytech com for updates to this manual and additional application information If you need technical or sales support please contact local...

Reviews: