11. OPERATION

The respirator system will turn off the turbo unit if the OFF button is pressed for more than 3 seconds.

The respirator system will cut off the entire circuit and switch to sleep mode if the turbo unit has turned off for more than 30
minutes. Pressing the ON button can activate the system.

The respirator system must be operated in the temperature range of -5°C to +55°C and relative humidity less than 90%RH.

12. TROUBLE SHOOTING

Problem

Probable cause

Action

Motor is stuck

Motor is damaged

Blower structure
failure caused by
outer force

Circuit failure

Check and remove
physical failure and
restart the system.
Return to dealer if
LCD still shows E01

Motor is damaged

Motor impeller rubs
blower shell

Circuit has ecessive
current

Check and remove
physical failure and
restart hte system.
Return to dealer if
LCD still shows E02

Low battery

Change the battery

Filter blocked

Tube blocked

Remove
obstruction
and/or change
the filter

Clean the tube

Battery high
temperature

Stop working and
rest

No air flow, no
alarm

No power

Battery contact
damaged

Charge the battery
and check battery
contact

Battery run time is
too short

Battery is not fully
charged

Filter is blocked

Battery is damaged

Charge the
battery

Remove
obstruction
and/or change
the filter

Change the
battery

Air supply to hood
smells unusual

Filter is broken

Tube is broken

ADF helmet is
broken

Leave current area
immediately

Change filter

Change tube

Change ADF
helmet

Supply insufficient
air to helmet

Breathing tube
broken off

Breathing tube
broken

Filter is blocked

Check the tube
connection
to hood and
respirator
system unit

Change
breathing tube

Remove
obstruction,
change filter

Summary of Contents for SPARX SINGLE

Page 1: ...SPARX SINGLE AIR INSTRUCTION MANUAL...

Page 2: ...ion is given If you are not sure about the concentration of pollution or about equipment performance ask the industrial safety engineer Warnings The respirator system devices should be used by well tr...

Page 3: ...2 The respirator protection system turbo unit filter waist belt 3 Lithium ion battery 4 The tube its anti fire cloth and both end fittings 5 The shoulder harness 6 The air flow tester 7 The lithium io...

Page 4: ...device was stored for more than 15 days When the battery is new or has been stored for more than 3 months charge it and discharge it at least twice in a row to reach the nominal rated charge capacity...

Page 5: ...system must be stored in a dry clean area in the temperature range of 10 C to 55 C and relative humidity less than 90 RH If the equipment is stored at temperature below 0 C the battery must be allowed...

Page 6: ...English 6 English 9 INSTALLING ON THE BELT Make sure the belt is securely fastened...

Page 7: ...airflow must be tested before using If the marble can t reach the minimum flow level do not use the system Change the filter or the battery and retest the air flow MUST TEST AIR FLOW ALARM TEST If th...

Page 8: ...has ecessive current Check and remove physical failure and restart hte system Return to dealer if LCD still shows E02 Low battery Change the battery Filter blocked Tube blocked Remove obstruction and...

Page 9: ...n Air Flow Rate and Filter Load LCD Display Air flow level and data Battery capacity Filter status Belt Size 35 2 5 x 51 2 5 in 900mm x 1300mm 14 MARKING EXPLANATION Powered filtering device EN 12941...

Page 10: ...13 02 111 Airduct with headgear 4 13 02 411 Inner visor clear lens 5 13 01 043 Lock sliders 6 13 03 101 Filter cover 7 13 03 104 Spark Arrestor 8 13 03 103 Pre filter 9 13 03 102 P3 level filter 10 13...

Reviews: