44
10. Calibration
Read Chapter 7 Before Use of
These Pages
10.3 Calibration Setting
10.3.1 Selecting the Calibration Type
Set the calibration type that will be used in Calibration Execution.
To set up “Span calibration using the built-in weight”,(UW only)
Select menu item
.
To set up “Calibration check using the built-in weight”,(UW only)
Select menu item
.
To set up “Span calibration using external weights”,
Select menu item
.*
To set up “Calibration check using external weights”,
Select menu item
.*
*Not applicable to a verified balance as a legal measuring instrument in the EU
10.3.2 PSC Fully-automatic Calibration (UW series only)
With the PSC function, span calibration is performed automatically using the built-in calibration weight
when the balance detects a temperature change that would affect weighing accuracy.
1
To turn ON the PSC function, Select menu item
.
2
To turn OFF the PSC function, Select menu
item .
1
2
3
4
5
6
Notes
• Blinking calibration symbol indicates an automatic calibration is about to
start.
• If PSC starts while the balance is in use, press the
[POWER]
key to abort
that cycle.
Note
Using a verified balance as a legal measuring instrument in the EU:
When PSC, fully-automatic span calibration, is not activated, operator must
carry out span calibration with the built-in weight (refer to
blinking of the Weight Symbol.