Chapter 5 Test and Material List
5.2 NIBP Calibration
Figure 5-1 NIBP Calibration
Calibration method:
Based on the precision of 50mmHg (6.7kPa), increase the pressure step by step. The
maximum error at any pressure point within the NIBP measurement range of the patient
monitor should be no more than ±3mmHg (±0.4kPa). Decrease the pressure step by step.
The maximum error at any pressure point within the NIBP measurement range of the patient
monitor should be no more than ±3mmHg (±0.4kPa).
5.3 IBP CALIBRATE
5.3.1 IBP Transducer Zero
Press the ZERO button on the IBP module to call up IBP PRESSURE ZERO menu as
shown below:
59