![Shenzhen PM-9000 Express Service Manual Download Page 51](http://html1.mh-extra.com/html/shenzhen/pm-9000-express/pm-9000-express_service-manual_1255785051.webp)
Chapter 3 Product Specifications
3.13 IBP Specifications
Number of channels
2
Pressure labels
ART, PA, CVP, RAP, LAP, ICP, P1 and P2
ART
0 to 300 mmHg
PA
– 6 to 120 mmHg
CVP/RAP/LAP/ICP
– 10 to 40 mmHg
Measurement range
P1/P2
– 50 to 300 mmHg
Resolution 1
mmHg
Precision
±2% or ±1mmHg, whichever is greater
Update period
1s
Pressure transducer
Sensitivity
5 uV/V/mmHg
Impedance range
300 to 3000
Ω