Shel lab SMO5CR-2 Installation And Operational Manual Download Page 15

15

 

Unit will heat up at all 

 

 

1/ verify that controller is asking for heat by looking for controller 
light 

– if controller light is not on, there is a problem with the 

controller 
2/check amperage 

– amperage should be virtually at maximum rated 

(data plate) amperage 
3/ do all controller functions work? 
4/ is the High Limit set high enough? 

– for diagnostics, should be 

fully clockwise with the pilot light never on 
5/ has the fuse/circuit breaker blown? 

Indicated chamber temperature 
unstable 

 

 

1/ ±0.1 may be normal 
2/ is blower motor working? 

– remove top panel and verify 

movement of blower motor in shadow box 
3/ is ambient room temperature radically changing 

– either door 

opening or room airflow from heaters or air conditioning? 

– stabilize 

ambient conditions 
4/ it may happen if exhaust stack is 100% open or if power exhaust is 
cycling 

– adjust stack to at least ¼ closed 

5/  sensor miss-located, damaged or wires may be damaged- check 
mounts for control and High Limit sensors, then trace wires between 
sensors and controls 
6/  calibration sensitivity 

– call Customer Service 

7 / high limit set too low 

– be sure that High Limit is more than 5 

degrees over desired set point; check if High Limit pilot is on 
continuously; turn controller knob completely clockwise to see if 
problem solved then follow instructions in Section 7.5 for correct 
setting  
8/ electrical noise 

– remove nearby sources of RFI including motors, 

arcing relays or radio transmitters 
9/ bad connection on temperature sensor or faulty sensor 

– check 

connectors for continuity and mechanical soundness while watching 
display for erratic behavior; check sensor and wiring for mechanical 
damage 
1 0 /   bad connections or faulty solid state relay 

– check connectors 

for mechanical soundness and look for corrosion around terminals 
or signs of arcing or other visible deterioration   

Will not maintain set point 

 

 

1 /  assure that set point is at least 5 degrees over ambient room 
temperature 

 

2/  see if ambient is fluctuating

 

Display and actual reference 
thermometer don’t match 

 

 

1/ calibration error 

– see Jumo users manual 

2/ temperature sensor failure 

– call Customer Service  

3/ controller failure 

– call Customer Service 

4/ allow at least two hours to stabilize 
5/ verify that reference thermometer is certified 
See above

 

Can't adjust set points or calibration 

 

 

1/ turn entire unit off and on to reset 
2/ if repeatedly happens, call Customer Service 

Calibrated at one temperature, but not 
at another 

 

 

This can be a normal condition when operating temperature varies 
widely. For maximum accuracy, calibration should be done at or as 
close to the set point temperature. 

MECHANICAL 

Door not sealing 

 

 

1/ check physical condition of gasket 
2/ assure that gasket is in original location 

Summary of Contents for SMO5CR-2

Page 1: ...Previously Designated as CR1 2 INSTALLATION AND OPERATIONAL MANUAL 01 2013 4861580 P O Box 627 300 N 26th Avenue Cornelius Oregon 97113 EMAIL tech Shellab com INTERNET http www Shellab com 1 800 322...

Page 2: ...NCE SECTION 9 0 TROUBLESHOOTING AND SERVICE SECTION 10 0 PARTS LIST UNIT SPECIFICATIONS SCHEMATICS This unit is a special purpose oven for professional industrial or educational use where the preparat...

Page 3: ...and enter your claim on the form supplied by the carrier 1 2 Inspect for concealed loss or damage on the unit itself both interior and exterior If necessary the carrier will arrange for official inspe...

Page 4: ...you should consult your manual for further description or discussion of a control or user item 2 2 Indicates Adjustable Temperature 2 3 Indicates AC Power ON 2 4 Indicates High Limit Safety 2 5 Indic...

Page 5: ...c Avoid direct sun fast moving air currents heating cooling ducts and high traffic areas Allow a minimum of 30 cm between the unit and walls or partitions which might obstruct free air flow 3 3 Liftin...

Page 6: ...own set point preventing sample loss or unit degradation Please note that it is not recommended that the unit be allowed to operate using only the High Limit to control temperature as temperature unif...

Page 7: ...ealt with according to local regulations Your dealer can provide you with a power exhaust which greatly helps under these applications 5 3 Do not operate near noxious fumes 5 4 Do not place sealed or...

Page 8: ...mbled and cleaned Make sure unit is unplugged and cooled down before disassembly When disassembling use an 8 32 nut driver for removing all the cap nuts Remove the bottom liner insert before removing...

Page 9: ...9 Figure 2...

Page 10: ...Once the down arrow has been pushed use the up and down arrow to achieve desired set point 3 After reaching desired set point the display should blink once to enter the value 4 To Auto Tune push the u...

Page 11: ...C in 30 minutes and control for 1 hour at 100 C Then you want to ramp up to 150 in 30 Minutes and control for 2 hours Then you want to cool down to start Set Point with Power Exhaust Outlet activated...

Page 12: ...and the display should read SENS This enters you in the Input Configuration Menu To navigate through the menu press the up and down arrows Press the up or down arrow repeatedly until you reach OFFS in...

Page 13: ...en cleaning and assure all volatile or flammable cleaners are evaporated and dry before reattaching the unit to the power supply 8 2 Storage If the oven is to be turned off it can be reactivated witho...

Page 14: ...irm with front panel lights that controller is calling for heat 7 high limit failure confirm with front panel lights that High Limit is operating correctly 8 wiring problem check all functions and com...

Page 15: ...problem solved then follow instructions in Section 7 5 for correct setting 8 electrical noise remove nearby sources of RFI including motors arcing relays or radio transmitters 9 bad connection on tem...

Page 16: ...other loads are on the wall circuit Unit will not turn on 1 check wall power source 2 check fuse circuit breaker on unit or in wall 3 see if unit is on e g fan or heater and just controller is off 4...

Page 17: ...inlet 2800503 Heating Element 9570831 Hepa filter 9990503 High Limit Controller 1750571 High Limit Indicator Pilot Lamp 4650553 Leveling Feet 2700512 Main Temperature Controller 1750803 ON OFF Switch...

Page 18: ...5 lbs Dimensions Exterior WxDxH in Interior WxDxH in SMO5CR 2 CR1 2 35 x 30 25 x 38 17 x 20 x 20 125 Capacity Cubic Feet SMO5CR 2 CR1 2 4 0 Temperature Range Uniformity Control SMO5CR 2 CR1 2 Amb 15 t...

Page 19: ...19 WIRE DIAGRAM...

Page 20: ...includes components not manufactured by Manufacturer is not covered by this warranty Any alteration or removal of the serial number on Manufacturer s parts will void this warranty Under no circumstanc...

Reviews: