The 

VLW601 

program module has 4 buttons (fig. 21) which allow to perform all operational, control and programming  

  instrument functions.

  In the configuration menus, is possible:

 

a)

 Submenus and parameters access; press 

 to select and press 

 to access.

 

b)

 Parameter options choice: Press 

 to select the option and press 

 to store the option.

   Press 

 to exit without storing

 

c)

 Configure the parameter values; in some parameters the configuration is done by setting a value (eg., in the 

  

SET DISTANCE 4mA

 parameter is possible to change the the corresponding distance value, in mm): press

 

 to  

  

select the digit to be modified (the digit is highlighted in 

inverse 

), press

 

 to change the highlighted digits    

  

number, press 

 to save the set value and exit automatically. Press 

 to exit without storing. 

    In the display top right, during the settings, there is always a number, eg. “1.2”. This number is the menu or    

    parameter index that’s displayed. The menu structure is represented on page 7 and on pages 9÷10.

OK

OK

BK

4321

mm

D

!

OK

BK

-  Configuration access

-  Options confirmation

-  Parameters values confirmation

-  Parameters values modification

OK

BK

- Exit configuration 

- Back to previous menu

- Eco map (from RUN mode)

-  Parameters values selection

-  Parameters scroll 

Fig.21

PTU5x

 

- Configuration and Quick Start

  

5. QUICK START MODE

From “RUN” mode press 

 

to access the Quick

 

Setup menu

  

OK

OK

BK

4321

mm

D

!

OK

OK

BK

SET DISTANCE 4mA           0

SET DISTANCE 20mA

MEDIUM

FILTER COEFFICIENT

BLIND DISTANCE

DISPLAY

OK

OK

BK

SET DISTANCE 4mA           0

SET DISTANCE 20mA

MEDIUM

FILTER COEFFICIENT

BLIND DISTANCE

DISPLAY

OK

OK

BK

FILTER COEFFICIENT

010

0.4

Select the parameters by 
moving the cursor with 

and confirm with

 

;

press 

 to exit

With the 

VLW601 

module is possible to access two configuration modes for the 

PTU50- 51-56 

setting:

QUICK START

 

-  Menu with easy access for quick basic parameters configuration.

To access: from “RUN” mode press 

 

to the quick setup menu mode access,

 

 to exit

ADVANCED CONFIGURATION 

- Full menu with access to all parameters, including functional parameters.

  

It is recommended to carefully read the complete documentation before accessing.

To access: from “RUN” mode, 

holding down 

 

, press

 

 

to the advanced configuration mode access,

 

 to exit

WARNING!

 

- The documentation provided with the 

PTU50-51-56

 contain the most 

frequently used indications. If it’s necessary refer to the full manual, it can be downlo

-

aded from our website 

www.sgm-lektra.com

 , in the products section.

Summary of Contents for PTU5 Series

Page 1: ...Certificate In conformity to the company and check procedures I certify that the equipment PTU Production and check date Serial n is conform to the technical requirements on Technical Data and it is...

Page 2: ...PTU51 and PTU56 units are compact sensors and have a via connector quick connection The IP68 protection makes them suitable for external applications with direct exposure to the weather including are...

Page 3: ...for 12m max PTU56 range where the sensor can not measure PTU50 1 5m PTU51 6m PTU56 12m D mm Lmax mm D mm Lmax mm 100 80m 125 240 125 240 150 300 150 300 Fig 10 Fig 11 Tab 2 2 3 Mounting precautions 2...

Page 4: ...illing and emptying In the programming menu to the PRODUCT parameter must select the LIQUID PIPE option see page 7 or 11 2 3 5 Agitators presence The level measurement is possible thanks to the Auto T...

Page 5: ...tighten the connector ring nut position the cable so that it forms a downward curve at the M20 output Fig 18 in this way the condensation and or rain water will tend to drip from the curve bottom Fig...

Page 6: ...ification OK BK Exit configuration Back to previous menu Eco map from RUN mode Parameters values selection Parameters scroll Fig 21 PTU5x Configuration and Quick Start 5 QUICK START MODE From RUN mode...

Page 7: ...U50 1 5m max 300mm PTU51 6m max 500mm PTU56 8m max 5 1 Struttura men di configurazione rapida 4mA distance 20mA distance 100 0 Level measurement PTU50 51 56 referent point Fig 22 5 1 1 SET DISTANCE 4m...

Page 8: ...II 49 20090 Rodano MI ITALY tel 39 0295328257 fax 39 0295328321 web www sgm lektra com e mail info sgm lektra com Pagesfrom9to20 fullmanual canbedownloadedfromourwebsitewww sgm lektra comintheproducts...

Reviews: