46
7 Adjustment menu
7.4
Setting the O
2
sensor
O
2
- zero point 0 %VOL
now release test gas
100 vol.%
CH
4
from the SPE VOL
wait for the display to settle at
a stable value and confirm the
adjustment with the on/off key
(
OK
appears in the LCD)
once this has happened, turn
off the test gas feed
pressing the
arrow-up key
brings you to the next display
O
2
- sensitivity 20.9 %VOL
now release
fresh air
from
the SPE VOL (cf. section 7.8:
Setting the test gas concen-
tration)
wait for the display to settle at
a stable value and confirm the
adjustment with the on/off key
(
OK
appears in the LCD)
pressing the
arrow-up key
brings you to the CO
2
-sensor
setting facility