Quick Start Guide
SERVO
PRO CHROMA/NANOCHROME/ULTRA Gas Analyser
22
04400003A_0
2.
Select a peak
Name
with the mouse.
3.
Click on the
Add “
peak name
”
calib values
button and a pop-up menu
will appear (see Figure 11).
4.
Enter
the number of points
that will be used to calibrate the system. A
minimum of two points is required: point 0.00 by default and the span
value. The
Area
value will be calculated by the analyser.
5.
For a chassis fitted with an Internal Dilution System, the values of
F1
(
Flow rate with Zero only
) and
F2
(
Flow rate with Zero and Span
) are
set during factory configuration.
6.
Select
OK
and the pop-up menu will disappear.
7.
For multiple peaks, repeat the above steps for each peak name
8.
For each peak to be calibrated, left click on the arrow in the combo box in
the
Span Gas
column to update the values
Start the calibration:
9.
Enable the
Span Calibration
with the
Enable (F3)
button.
10.
In the
Span
column select
Yes
for the peaks that require calibration
11.
Select
All peaks
or
Selected peaks
by changing the slider state. When
Selected peaks
is selected, only peaks that have a
YES
in the
Select
column of the grid
will be calibrated.
12.
Click on
Start
(F6)
button to start calibration
13.
After three cycles with consistent peak values (within 3% FS) press the
RE-SPAN
button to accept the calibration. A pop-up box question will
appear
Are you sure you want to re-span?
Select
YES
to accept the
calibration. The
Result
column will be updated with the value set in the
Span Gas
column, and the
Response Factor
column will be re-calculated.
14.
When the
Span Calibration
is completed,
YES
is displayed in the
Span
Done
column of the grid.
15.
Repeat the above if there are other peaks to be calibrated.
16.
Disable
Span
Calibration
with the
Disable (F3)
button (one button
performs dual functions Enable or Disable, depending on the active mode).