Quick Start Guide
SERVO
PRO CHROMA/NANOCHROME/ULTRA Gas Analyser
15
04400003A_0
The
Range
cannot be changed during a peak integration. When
Automatic
ranging is selected, range selection will be automatically
based on the result of the last peak calculations.
If the peak selected is configured to ppb and the measurement
switches into Range 1, ppt measurement will be seen in Peak table.
If the measurement switches into Range 2, ppb measurement will be
seen in Peak table.
Note that when Threshold calibration is enabled,
“
Warning: Threshold
Calibration enabled
”
will be displayed in the chromatogram display with
Enhanced View On.
Note that when Span calibration is enabled,
“
Warning: Span Calibration
enabled
”
will be displayed in the chromatogram display with Enhanced View Off.
During both Threshold and Span calibration the
Enhanced View
radio control
will be disabled.
The colour displayed in the
Alarm column (Al)
of the peak grid indicates the
alarm status for each peak. The following table shows the meaning of each colour.
Table 1 : Peak Grid - Alarm Status display colours
1.6.9 Save (F6)
Trending data can be saved by pressing the
Save
button at the end of a cycle.
When the
Save
button is pressed, a dialog box appears and asks for a filename to
be entered. Load this file from the
ANALYSE CHROMATOGRAM MENU
.
The button is enabled only when no cycle is in progress.
Delete trending data that is no longer required with the
Delete
button
of the
ANALYSE CHROMATOGRAM
menu.
Colour
Alarm Status
White
No alarm
Yellow
Alarm 1
Red
Alarm 2