01440D Light Industrial Gas Analyser Service Manual.
2.2
2.2
Sample condition requirements
Table 2.1
Requirements for Sample Conditions
Dew Point
At least 5°C below ambient temperature.
Temperature
Nominally at ambient temperature.
Inlet / Outlet
connections
Note: do not
restrict analyser
vent
01440D1STD
6.4mm (1/4”) OD tube.
0144
BPR
01440D1FTX
3.2mm (1/8”) OD tube.
Inlet pressure
01440D1STD
1 to 10psig (7 to 70kPag).
0144
BPR
17 to 20psia (120 to 140kPaa).
01440D1FTX
Typically 0.3kPag (30mmWG) at 200ml/
min.
Flow Rate
01440D1STD
1 to 6l/min.
0144
BPR
1 to 2l/min.
01440D1FTX
USER LIMITED to 250ml/min. MAXIMUM.
Particulates
01440D1STD
<3µm (micro meters), an internal,
replaceable, 0.6µm filter is fitted as
standard.
0144
BPR
01440D1FTX
The user must fit an external filter of
0.6µm.
Condition
01440D1STD
Must not be flammable, toxic or corrosive,
as well as being clean and oil free.
0144
BPR
01440D1FTX
Must not be corrosive, as well as being
clean and oil free.