7
7
3. INSTALLATION
3-1. Place of Installation:
Install SEPAREL in accordance with conditions specified in the specification sheet,
●
No exposure to direct sunlight
●
No physical shock or shaking
●
No contact with dust, moisture, corrosive gas, or liquid
●
Easy access for maintenance, inspection, repair, and replacement
3-2. Method of Installation:
●
Set the vacuum port downward to smoothly purge liquid derived from vapor.
●
If you fix SEPAREL with a U-band, make sure not to loosen the fitting. Do not apply too
much pressure when fixing SEPAREL with a U-band as too much pressure may damage
the module. Putting a cushion between SEPAREL and U-band may prevent damage to
the module.
3-3. Connection of the SEPAREL Module:
●
Clean pipe/tube before connecting to prevent/remove dust, rust, oil, etc.
●
Do not apply too much pressure to the connection.
●
Do not shake or shock the SEPAREL module. Shaking or shocking SEPAREL
may damage the module even if it is a light impact.
●
Because the housing of the SEPAREL module is composed of PP, polypropylene,
excess pressure to the connector may cause damage to the connecting port. Excess
pressure to connecting port may shave thread and cause resin particles to appear in the
liquid side.
●
Do not touch hollow fibers.
●
When you add a connector, adjust the insertion depth and not the screw part to touch
hollow fibers. If the screw part touches the hollow fibers, the hollow fibers may become
damaged.
●
Depending on the type of connector, recommended conditions of connector insertion
may differ. For detailed conditions, including torque value, contact us.
●
The liquid feeding pump is recommended to be placed in front of the SEPAREL module.
●
When you uninstall the SEPAREL module, be careful to not damage it.
Application Materials Product Division
INSTRUCTION MANUAL