background image

                                                                                                                                                                           

ST805C PON Power Meter 

 

"►","◄",  and  then  press "OK"  to  save  or  press  "OFF"  to 
exit. 

Exit automatically without operation in 10 seconds. 

5.5 Auto power off 

Under main interface, short press "OFF" to switch between 

auto power off and manual power off, icon displays when 

auto power off; default auto power off without operation in 

10 minutes. 

5.6 Record browsing and management 

Under  main  interface,  long  press  "◄"  to  enter  record 
browse interface, then you can press 

"▲", "▼" to choose 

record. Long press "OK" to delete all records. 

Exit automatically without operation in 10 seconds. 

5.7 Battery under voltage 

Once the battery voltage is below 6.7V (for Lithium battery), 

the indicator will be 

, meter will start alarm and power off, 

please  charge  meter  (about  4-6  hours)  in  time,  charge 

indicator  turns  from  red  to  green,  it  means  battery  is  full 

charged. 

Summary of Contents for ST805C

Page 1: ...ST805C PON Power Meter User Manual V 1 1 Shandong Senter Electronic Co Ltd ...

Page 2: ...d Technical specification Function 3 4 Configuration 6 5 Operation Instruction 8 5 1 ON OFF 8 5 2 Relative value differential loss 8 5 3 Threshold setup 9 5 4 Time setting 9 5 5 Auto power off 10 5 6 Record browsing and management 10 5 7 Battery under voltage 10 6 Maintenance 10 ...

Page 3: ...an process pass fail analysis conveniently through multiple users adjustable threshold of every wavelength It adopts 32 digits CPU with low power consumption so the functions are more powerful and rapid More conveniently measurement for field and equipment site owes to friendly operation interface It s a useful site test tool for the engineers of PON network and maintenance operators 2 Main Featur...

Page 4: ...test 3 wavelengths power of PON system synchronously 1490nm 1550nm 1310nm Can test the burst mode signal of 1310nm Threshold value setting function Data storage function Auto backlight off function Display the voltage of battery low voltage power off Real time clock display Technical specifications Standard wavelengths 1310 upstream 1490 downstream 1550 downstream Pass zone nm 1260 1360 1470 1505 ...

Page 5: ...inuous working 18 hours for Lithium battery about 18 hours for dry battery too but different for different brands Working temperature 10 60 Storage Temperature 25 70 Dimensions mm 200 90 43 Weight g about 330 1490nm dB Isolation 1550nm dB 40 40 Accuracy Uncertainty dB 0 5 Polarization Dependent Loss dB 0 25 Linearity dB 0 1 Through Insertion Loss dB 1 5 Resolution 0 01dB Unit dBm xW ...

Page 6: ...ST805C PON Power Meter 5 4 Configuration Figure 1 Front view ...

Page 7: ...ST805C PON Power Meter 6 Figure 2 Top View ONT interface Connect with ONT equipment through pigtail when testing OLT interface Connect with OLT equipment through pigtail when testing Figure 3 Side View ...

Page 8: ... for 3 seconds is long press and press 1 second is short press similar hereinafter OFF Long press OFF for 2 seconds meter powers off Under main interface after power on Press F2 to process unit conversion dBm or xW Press F1 Record save prompt will shine press OK to complete the record storage exit save state automatically without operation in 10 seconds 99 records could be stored 5 2 Relative valu...

Page 9: ... to enter threshold edit state You can use edit threshold value and use select value location press OK to save and exit press OFF to exit without save When browsing threshold value in management interface short press OK to enable current threshold value then you can press OFF to exit Automatically exit test interface without operation in 10 seconds Under main interface long press OK button to dele...

Page 10: ...ation in 10 minutes 5 6 Record browsing and management Under main interface long press to enter record browse interface then you can press to choose record Long press OK to delete all records Exit automatically without operation in 10 seconds 5 7 Battery under voltage Once the battery voltage is below 6 7V for Lithium battery the indicator will be meter will start alarm and power off please charge...

Page 11: ...d surface or else it will damage the sensor s head face 2 Change the adapter carefully keep the spared adapters with sealing tight to avoid of touching dirt 3 Please cover the dust cap timely if meter is unused keep the head face clean don t expose the sensor in air for long time or inaccuracy of measurement will be caused if sensor touches dirt 4 Clean the sensor s head face regularly ...

Page 12: ... Meter 11 Shandong Senter Electronic Co Ltd No 18 Liu Yishan Road New Hi tech Area Zibo City Shandong Province China 255086 Tel 86 533 3589257 Fax 86 533 3587522 3589257 Email export senter com cn http en senter com cn ...