1257 Salt Springs Rd.
Youngstown, OH 44509-1612
Phone: (800) 239 -1226
Fax: (801) 697-0825
People Counter with Data Recorder
Page 2 of 4
The count recorder can be mounted or set in any convenient indoor location. Place near the sensor for
autonomous operation or near your computer for a permanent PC connection. For more system layout
suggestions refer to the count recorder manual.
3. Wiring Components
Connect the wires from the sensor, power supply, count recorder, and chime (optional) to the terminal
strip according to the wiring diagram, while adhering to the wiring codes established for your state and
locality. Excess wire length can be trimmed. Please follow the included instructions with the count
recorder along with the wiring guide in figure 1.
*
Optional component
4. Aligning Sensor
After the sensor and count recorder are mounted and wired, connect the power supply to a 120V AC
outlet and take note of the lights on the sensor.
If the sensors sensitivity and alignment are correct NO LIGHTS should be on when nothing is in front of
the sensor. The lights should change from NO lights to RED or YELLOW when a person enters the
sensing area (RED indicates marginal sensing) and the count will be incremented by +1. Adjustment
may be necessary if lights are on when nothing is in front of the sensor. When finished, secure the
sensor in order to maintain alignment.
Using the supplied adjustment tool, sensitivity of each sensor may be increased or decreased as
needed by turning the small adjustment screw on the side of the sensors (clockwise to increase or
counter-clockwise to decrease sensitivity).
NOTE: Testing sensitivity of each sensor should be done with a person wearing dark clothing.
5. Your counting system is ready for operation
NOTE: Please refer to the Count Recorder manual for further information.
Refer to Figure 1