RK/WK
Read/Write Known-Load Calibration
Point
Purpose
Reads or writes the channel’s "
KKKKNNNNOOOOWWWWNNNN POINT x
POINT x
POINT x
POINT x////yyyy
” menu items
Usage
"
#
aacc
RK
pp
↵
" to read, "
#
aacc
WK
ppn
↵
" to write.
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
pp
is the parameter defined below.
n
is the known-load calibration point in engineering units.
↵
is the ‘carriage return’ character (ASCII decimal 13).
Parameter
"
00
” accesses the "
KKKKNNNNOOOOWWWWNNNN POINT 1
POINT 1
POINT 1
POINT 1////5555
” menu item.
"
01
” accesses the "
KKKKNNNNOOOOWWWWNNNN POINT 2
POINT 2
POINT 2
POINT 2////5555
” menu item.
"
02
” accesses the "
KKKKNNNNOOOOWWWWNNNN POINT 3
POINT 3
POINT 3
POINT 3////5555
” menu item.
"
03
” accesses the "
KKKKNNNNOOOOWWWWNNNN POINT 4
POINT 4
POINT 4
POINT 4////5555
” menu item.
"
04
” accesses the "
KKKKNNNNOOOOWWWWNNNN POINT 5
POINT 5
POINT 5
POINT 5////5555
” menu item.
Example
"
#0001RK01
↵
”
Reply
When writing: "
OK
" or "
ERROR
”.
When reading: an ASCII-floating-point value.
Changing this value has no effect until the channel is re-cali-
brated to the RTD with the Known-Load Calibration method.
Summary of Contents for SC1000
Page 7: ...Additional Channels for SC Series Instrumentation page 7 INTRODUCTION 1 ...
Page 8: ...page 8 008 0608 01 ...
Page 31: ...Additional Channels for SC Series Instrumentation page 31 RTD INPUT CHANNEL 2 ...
Page 32: ...page 32 008 0608 01 ...
Page 53: ...Additional Channels for SC Series Instrumentation page 53 FREQUENCY INPUT CHANNEL 3 ...
Page 54: ...page 54 008 0608 01 ...
Page 58: ...page 58 008 0608 01 ...
Page 60: ...page 60 008 0608 01 ...