page 12 008-0608-01
2.4 Calibration
A listing of all menu items for this channel is given in “Setup Menu Reference” on
page 55.
A customer-supplied RTD simulator/calibrator is required for re-calibration.
Although not recommended, field re-calibration of the channel for operation
with RTDs with outputs other than the Pt100 DIN-43760 / IEC751 standards
can be made; this may result in reduced accuracy and resolution.
Step 1: Wire the RTD simulator/calibrator to the channel’s connector.
Step 2: Enter the
CCCCAAAALLLLIB
IB
IB
IBRRRRAAAATTTTIIIIOOOONNNN DAT
DAT
DAT
DATAAAA
.
If you wish to re-calibrate the instrument using a unit-of-measure other than
degrees Celsius, you can use the
CCCCAL
AL
AL
ALIIIIBBBBRRRRAAAATTTTIIIIOOOON U
N U
N U
N UNNNNIIIITTTTSSSS
menu item to change the
units-of-measure that the instrument uses for calibration.
Step 3: Perform the calibration.
Using the
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTEEEE
menu item starts the calibration process. You will be
prompted to use your RTD simulator/calibrator to apply simulated RTD signals
to the channel as required.