Cambridge Sensotec Rapidox 1100ZR-H-OL O
2
Analyser Instruction Manual
Version 1.2
42
Last printed 22/07/2015 18:02:00
USB-Serial port adapter which does not have an FTDI chipset then you may
encounter difficulties. Contact us for advice!
Q:
I selected ‘Write to the Analyser’ and now the Rapidox is way off
calibration.
A:
Only select ‘Write to the Analyser’ once all the values are correctly entered
in the boxes on the right. The best procedure is to ‘Read Analyser
Configuration’ and then press ‘Copy’ so the values in the boxes are the same
as those stored in the Rapidox memory. You can then edit these values and
then select ‘Write to the Analyser’ when finished.
Q:
How do I get back to the factory calibration for my instrument.
A:
Select ‘Load Configuration File’ from the File menu and look for the rxc file
on your memory stick. Select this file and then select ‘Write to the
Analyser’. This will load the factory calibration back into the Rapidox.
Alternatively you can load a standard default configuration any time using
the ‘Load Default’ option on the File menu. This configuration will not be
perfect for your analyser but will be good enough to get you started again.
10.1
Warranty
The Rapidox analyser has been carefully tested and inspected before shipment
and is guaranteed to be free from defective materials and workmanship for a
period of twelve months from date of purchase. The sensor head is replaceable
and has a life expectancy in excess of 17,500 hours. However, if the analysis gas
contains corrosive gases or large quantities of particulates, sensor life may be
shortened. In the case of the latter, it is normally possible to insert a filter prior to
the sensor head to remove the particulate material.
10.2
Conditions of Warranty:
1) This warranty is in addition to and does not affect any statutory rights of
consumer purchasers. This warranty is valid worldwide on a “back to base basis”.
2) This warranty covers breakdowns due to design or manufacturing faults; it does
not apply to damage, however caused, wear and tear, neglect, unauthorised
adjustment or repair, or any items of limited natural life.
3) In the event of failure, please take the following action:
a] Refer to the “Troubleshooting” section of your instruction manual to
identify and possibly correct the problem.
b] If the fault cannot be resolved, please contact the Cambridge Sensotec
service and repair centre at the address given on the cover of the manual.