Cambridge Sensotec Rapidox 1100ZR-H-OL O
2
Analyser Instruction Manual
Version 1.2
40
Last printed 22/07/2015 18:02:00
9.24
DDE Protocols
For the Rapidox software to communicate with the main oven software using DDE
please refer to the following list of possible links that are currently used:
PFCControl state: 0 = Off, 1 = Monitor Only, 2 = On
"RAPIDOX|Main!PFCControl"
Remote Data Acquisition: "RUN", "PAUSE", "RESUME", "STOP"
"RAPIDOX|Main!ACQControl"
Desired O2 level in PPM: High and Low window set as PPM/- 20ppm.
Included for backwards compatibility
"RAPIDOX|Main!PPM_level"
Desired O2 level window in PPM
"RAPIDOX|Main!HighO2SetPoint"
"RAPIDOX|Main!LowO2SetPoint"
Current O2 level in PPM
"RAPIDOX|Main!lbl_PPM_O2"
Set to -1 when not measuring, see Error state for reason, -4 to indicate sensor fault
and -5 to indicate under range error.
Current O2 Level in %
"RAPIDOX|Main!lbl_PCNT_O2"
Set to -1 when not measuring, see Error state for reason, -4 to indicate sensor fault
and -5 to indicate under range error.
Error state indication: Error ID:Description
Error ID: -5 = O2 Range Error, -4 = Sensor Fault, -3 = No available com ports, -2 =
Communication problem, -1 = No analyser found, 0 = Analyser waiting, 1 = No Error
"RAPIDOX|Main!Error"
Not used, included only for backward compatibility
"RAPIDOX|Main!lbl_ReadyIn_XXmin"
Set points for SW alarms in PPM.
"RAPIDOX|Main!Alarm1"
"RAPIDOX|Main!Alarm2"
"RAPIDOX|Main!Alarm1_Ch2"
"RAPIDOX|Main!Alarm2_Ch2"
"RAPIDOX|Main!Alarm1_Ch3"
"RAPIDOX|Main!Alarm2_Ch3"
Current alarm states: -1 = Disabled, 0 = No Alarm, 1 = Alarm
"RAPIDOX|Main!Alm1_Ch1_State"
"RAPIDOX|Main!Alm2_Ch1_State"
"RAPIDOX|Main!Alm1_Ch2_State"
"RAPIDOX|Main!Alm2_Ch2_State"