Page 36
Oryx DTS User Manual v4
Once all measurement parameters have been set, the user is ready to upload the
configuration to an Oryx. Select the Oryx required and click “Upload”. The user is
prompted to correct any errors in the configuration and warnings (such as Channel
X not activated) are displayed. Errors will need to be corrected by clicking “Edit
values” while warnings are for information only.
NOTE: Clicking “UPLOAD” will upload a new configuration to the Oryx and will
DELETE ALL MEASURMENTS on-board the Oryx at that time. (Any data already
downloaded to the PC is unaffected).
The user is prompted to select the start time of the measurement cycle (see Figure
23).
NOTE: Ensure the Oryx date-time has been synchronised to the base station. Date
& time entered should be Oryx time.
NOTE: If selecting to start measurements as soon as possible, there may be a
delay of up to one repetition time before commencement of the first measurement.
Figure 23 - Selecting the start time of measurement cycle
MEASUREMENT CONFIG WIZARD
The “Measurement config wizard” option allows the user to be guided step-by-step
through the DTS setup, ensuring the calibration parameters are optimised for the