Operator Manual
Page
33
of
40
P/N 1580281
SEC3120 Digital Transmitter
Sensor Electronics Corporation
Rev 10, 20150707
Events such as a 5% gas level change, alarm state change, sensor warm-up, calibration, system power
on, sensor fault or parameter changes are logged and stored. Data from normal operation is NOT recorded
when gas levels do not change beyond a 5% band. Here is the data log format (SEC Filename:
“SEC3100.TXT”):
LOG FILE EVENT ENTRY FORMAT:
tt,mm/dd/yyyy,HH:MM:SS,ID- vv
Where:
tt
= type, 00 - 99
mm
= month, 1 - 12
dd
= day, 1 - 31
yyyy
= year, 2000 - 2099
HH
= hours, 0 - 23
MM
= minutes, 0 - 59
SS
= seconds, 0 - 59
ID-
= Sensor number (S1, S2 or BB for both)
vv
= variable data depending on tt
Log Entry Type Table (tt):
Log Entry Type
Code (tt)
Description
00
Boot/Power up- 3120 Unit Information Event
01
Sensor Warm-up: New Sensor/Sensor removed and replaced Event
02
Parameter Changes Made Event
03
Alarm High-Level Triggered Event
04
Alarm Mid-Level Triggered Event
05
Alarm Low-Level Triggered Event
06
Change in Gas Level > 5% of Sensor Range Event
07
Sensor Fault/Missing/Not Identical Detected Event
08
Sensor Calibrated Event
09
Flash Cleared/Restarted Event
10
Self-Test Initiated by Operator Event
11
Self-Test Concluded by Operator Event
12
Self-Test Aborted by Operator Event
13
Sensor Cell Warning Event
14
3120 Unit Role Change Event
99
Flash Card Re-inserted Event
Log Entry Data Format For Each Type Table (vv):
Log Entry Type
Code (tt)
Log Entry Event Name
Log Variable Data Format (vv)
00
3120 Boot/Power Up Event
Preamble,BB- US,FWVerMaj.Min.Rev
01
Sensor Warm-up Event
Preamble,ID- US,FWVerMaj.Min.Rev
02
Parameters Changed Event
Preamble,ID- NID,ZID,SSN,STP,CD,CV,RNG,LOW,MID,HI,GU,GN
03
Alarm High-Level Triggered Event
Preamble,ID- Gas Value Float
04
Alarm Mid-Level Triggered Event
Preamble,ID- Gas Value Float
05
Alarm Low-Level Triggered Event
Preamble, ID- Gas Value Float
06
Change in Gas Level Event
Preamble, ID- Gas Value Float