Calibration Menu
Enter
Using the Up and Down arrows allows the operator to move the cursor (*) to select a function.
To Zero the sensor apply clean air (N2 for an oxygen sensor) and select enter. The following will be
displayed.
Then the following will be displayed.
Once complete the following will be displayed. The sensor has been successfully zeroed.
Arrow down to CAL. VAL to verify the span gas calibration value matches the value of the span gas
calibration on hand. If not, select Enter and the following screen will appear.
Using the Up and Down arrows will allow the operator to change the calibration gas value of the sensor to
match the calibration gas used to span the sensor. Once the correct value is displayed select Enter and the
sensor will be uploaded with the new calibration gas value.
* CALIBRATION
ALARM
RELAY
ZERO
SPAN
* CAL. VAL 5 *
EXIT
* ZERO
0
SPAN
CAL. VAL
EXIT
CALIBRATION PROCESS
------- WAIT ----------
CALIBRATION PROCESS
------- DONE ----------
* ZERO
0
SPAN
CAL. VAL
EXIT