19
Server Address: sensit-test-hub.azure-devices.net
Access Point Name: zipitwireless.com.attz
GPS Mode: Interval
Sensor Heater: Enabled
Temp Set: 15
Dew Point Control: Disabled
Humidity Threshold: 80
Temp Set: 15
PID1 Offset Value (ppb): 0.00
Sample Trigger Value (ppb): 500
Trigger Average Time (sec): 10
Sample Collection Time (sec): 30
GPS:
“ENABLE” or “DISABLE” the GPS. Depending on the cellular data protocol
and output mode, the user may be presented with additional options. If cellular
output is disabled or if PERIODIC protocol is selected the GPS mode will default to
SINGLE (see below). If cellular output is enabled
and
MQTT or HTTP protocol are
enabled the user can select between the following GPS settings:
“CONSTANT”: GPS is always on and fixing position (Highest power draw)
“INTERVAL”: GPS turns on at and interval determined by “RATIO”x”ODR”
“SINGLE”: GPS turns on at power up and remains off until system is reset
HEAT:
Adjust “CONSTANT” or “HUMID” heat settings. The “CONSTANT” setting
means that the sensor heater is on at all times if enabled. The “HUMID” setting
means that the sensor heater is on only if humidity exceeds a predefined
threshold. If either option is enabled the user will be prompted to adjust the
heater setpoints and for the “HUMID” option the user will be prompted to adjust
the humidity set point. Both options could be enabled simultaneously to produce
different heater setpoints depending on the humidity level. “CONSTANT” heat is
enabled by default at 15°C above ambient.
OFFSET:
Adjust the baseline of the PID sensor up or down by a constant value.
OUTPUT:
This menu contains settings that modify the output characteristics of
the sensor.