Sensidyne Gilian AIRCON-2 Operation & Service Manual Download Page 7

7

Gilian

®

 

AIRCON-2 HIGH VOLUME AIR SAMPLER

Sensidyne Document No. F-PRO-3100 (Rev D)

READ AND UNDERSTAND ALL WARNINGS BEFORE USE

Read and understand 

ALL 

warnings before using this product. Failure to read, understand, and comply with

ALL 

warnings could result in property damage, severe personal injury, or death.

Read and understand 

ALL 

applicable Federal, State, and Local environmental health and safety laws and

regulations, including OSHA. Ensure complete compliance with 

ALL 

applicable laws and regulations before

and during use of this product.

UNDER NO CIRCUMSTANCES 

should this product be used except by qualified, trained, technically

competent personnel and not until the warnings, 

Operation and Service Manual

, labels, and other literature

accompanying this product have been read and understood.

The 

Operation and Service Manual must be read and understood by each user before operating this product

or using its accessories, in order to ensure proper and safe use and installation of this product and to ensure
familiarity with the proper treatment and safety procedures in the event of an accident.

Caution

:  Risk of electrical shock.

Warning

:  Line cord must be removed to remove all power from equipment.

DO NOT 

remove, cover, or alter any label or tag on this product, its accessories, or related products.

DO NOT 

operate this product should it malfunction or require repair. Operation of a malfunctioning product,

or a product requiring repair may result in serious personal injury or death. 

DO NOT 

attempt to repair or

modify the instrument, except as specified in the 

Operation and Service Manual

. Contact the Sensidyne

Service Department to arrange for a Returned Material Authorization (RMA).

Use 

ONLY 

genuine Sensidyne

®

 replacement parts when performing any maintenance procedures described

in this manual Specifically the DC version of the AirCon-2 is designed for use with AirCon-2 battery packs,
and should not be used in conjunction with any other equipment. Rechargeable battery packs are designed
to be charged with the AirCon-2 Power Module and in strict accordance with the charging instructions
contained in this manual. 

Failure to follow these warnings may seriously impair instrument performance

.

Repair or alteration of the product beyond the scope of these maintenance instructions, or by anyone other
than a certified Sensidyne

®

 serviceman, could cause the product to fail to perform as designed, and persons

who rely on this product for their safety could sustain severe personal injury or death.

DO NOT 

operate unit in excessive chemical or water vapor atmospheres. Failure to follow instructions may

cause permanent damage to the equipment.

NEVER 

block the cooling air inlets (located at the top/back of unit) when the unit is operating.

NEVER 

block any of the venting holes (located on the sides of the case) when the unit is operating.

ALWAYS 

place the unit on a secure and level surface when operating.

ALWAYS 

operate unit within the intrinsically safe environmental regime specified (non-hazardous loca-

tions).

WARNINGS !

Summary of Contents for Gilian AIRCON-2

Page 1: ...0550 FAX www sensidyne com Revision D Document No F PRO 3100 Gilian AIRCON 2 HIGH VOLUME AIR SAMPLER OPERATION SERVICE MANUAL CYCLE MODE LOW BATTERY FAULT ERROR RUN HOLD PROGRAMMING SINGLE MODE DELAY...

Page 2: ...ality performance and accuracy of your Sensidyne product Be assured of our complete and prompt answers to any questions you may have about this product since we are totally committed to providing you...

Page 3: ...gistered trademarks of Sensidyne Inc These trademarks are protected through use and registration in the United States The trademarks and servicemarks used in this document are the property of their re...

Page 4: ...10 1 2 4 Power Module 10 1 3 Controls Indicators 12 1 3 1 On Off Switch 12 1 3 2 Flow Adjustment 12 1 3 3 LCD Display Program Model 12 1 3 4 Touch Pad Buttons Program Model 12 1 4 Standard Parts 12 1...

Page 5: ...Basic Operation 21 3 2 1 Single Mode All Models 21 3 2 2 Cycle Mode Operation 21 3 3 Faults 24 3 3 1 Fault Function Program Model 24 3 3 2 Correcting Fault Conditions 24 3 3 3 Low Battery Indication...

Page 6: ...Hose 11 1 2 Front Panel Program Model 13 1 3 Liquid Crystal Display Program Model 14 1 4 AirCon 2 Accessory Configurations 15 2 1 DC Sampler Set Up 18 2 2 Power Module Set Up Single Pack 19 2 3 Power...

Page 7: ...product requiring repair may result in serious personal injury or death DO NOT attempt to repair or modify the instrument except as specified in the Operation and Service Manual Contact the Sensidyne...

Page 8: ...on 2 AC models include a Power Cord AirCon 2 DC models require a battery pack sold separately Air Hose Tubing Hose Support Tip Sampling Mast Operation and Service Manual Registration Card Warranty Car...

Page 9: ...which the air hose support tip is connected The media typically a filter cassette is inserted into the end of the sampling hose The other end of the air hose is connected to the Air Inlet located at...

Page 10: ...adapter for the voltage receptacle IMPORTANT You should place a label or sticker on the unit to show that the voltage has been changed NOTE Changing factory set 115 VAC units to 230 VAC negates CE ce...

Page 11: ...lug On Off Switch LCD Display ACCEPT Key 1 2 3 4 5 6 7 8 9 10 11 12 PROGRAM Key TIME PRESS Key RUN HOLD Key Inlet Air Boss Suction Flow Adjust Knob Flowmeter Sampling Mast Air Hose Hose Support Sampli...

Page 12: ...ttons are explained as follows ACCEPT This is used to switch the unit into the prepro grammed cycle modes or as confirmation in pro gramming mode to accept data into memory PROG Program This button is...

Page 13: ...F PRO 3100 Rev D Figure 1 2 Front Panel Program Model PROG RUN HOLD TIME PRESS ACCEPT OFF ON Made in USA Gilian High Volume Air Sampler CYCLE MODE LOW BATTERY FAULT ERROR RUN HOLD PROGRAMMING SINGLE...

Page 14: ...HIGH VOLUME AIR SAMPLER 14 Sensidyne Document No F PRO 3100 Rev D Figure 1 3 Liquid Crystal display Program Model MODE LOW BATTERY FAULT ERROR RUN HOLD PROGRAMMING SINGLE MODE DELAY TIME RUN TIME HOL...

Page 15: ...PRESSURE PROG TIME PRESS ACCEPT RUN HOLD CYCLE MODE LOW BATTERY FAULT ERROR RUN HOLD PROGRAMMING SINGLE MODE DELAY TIME RUN TIME HOLD TIME CYCLES PRESSURE 4 Hour Battery Pack mounted on the Power Mod...

Page 16: ...and extend the next section of the mast Tighten the second largest collar 4 Repeat this procedure until the mast is extended to its full or required height 5 Be sure to tighten each locking collar bef...

Page 17: ...ry packs and or power modules Disconnecting the DC Unit 1 Disengage the latch hook 2 Rotate the sampler about guide pin until the back of the sampler disengages the battery pack or power module 3 Dise...

Page 18: ...PRELIMINARY Gilian AIRCON 2 HIGH VOLUME AIR SAMPLER 18 Sensidyne Document No F PRO 3100 Rev D Figure 2 1 DC Sampler Set Up ELECTRICAL CONNECTION LATCH HOOK GUIDE PIN DC PACK...

Page 19: ...Gilian AIRCON 2 HIGH VOLUME AIR SAMPLER Sensidyne Document No F PRO 3100 Rev D Figure 2 2 Power Module Set Up Single Pack ELECTRICAL CONNECTION GUIDE PIN LATCH HOOK POWER MODULE POWER MODULE BATTERY P...

Page 20: ...Y Gilian AIRCON 2 HIGH VOLUME AIR SAMPLER 20 Sensidyne Document No F PRO 3100 Rev D Figure 2 3 Power Module Set Up Multiple Packs ELECTRICAL CONNECTION GUIDE PIN LATCH HOOK POWER MODULE BATTERY PACK B...

Page 21: ...When a fault occurs the hold function keeps the display visible so long as adequate battery power is available If the battery level drops further the display is erased How ever the unit stores the los...

Page 22: ...wmeter No display Sampler will run continuously at the set air flow rate until it is turned OFF BASIC MODEL PROGRAM MODEL ON OFF FLOW ADJ KNOB ACCEPT Option 1 RUN Option 2 PROG ACCEPT Note Switch powe...

Page 23: ...Press to take unit out of Run Mode and put into Hold Mode Motor turnsOFF clock stops counting colon stops flashing Display indicates SINGLE MODE HOLD Press at any time during RUN operation to view cu...

Page 24: ...r kinks c Re select a program number 1 3 refer to Table 3 4 for details 3 3 3 Low Battery Indication DC model When a DC battery pack is running out of charge the AirCon 2 indicates this condition on t...

Page 25: ...e display during the third RUN phase represents the total accumulated run time since the program started When the third RUN phase is com pleted the program concludes and the unit goes into a HOLD phas...

Page 26: ...is Off and the display counts down the time remaining until start up Press this button at any time during RUN operation to view current the Run Time or Back Pressure readings Display toggles between...

Page 27: ...s You must select another program number Step Switch Button Action Display Example FAULT PROGRAMMING HOLD SINGLE MODE RUN TIME PROGRAMMING 1 2 3 4 5 ON OFF FLOW ADJ KNOB ACCEPT PROG ACCEPT Switch to O...

Page 28: ...s and hold until hour increment 00 99 is reached Press to lock in DELAY TIME Hours and advance to set up for DELAY TIME Minutes Begin set up for DELAY TIME Minutes Minute position is flashing Press an...

Page 29: ...EPT PROG ACCEPT Begin set up for RUN TIME Hours Hour position is flashing Press and hold until hour increment 00 99 is reached You must enter a value greater than zero for unit to operate Press to loc...

Page 30: ...s flashing If you do not want to enter HOLD TIME data press ACCEPT twice to advance display to RUN CYCLES or continue to step 3 Press and hold until hour increment 00 99 is reached Press to lock in HO...

Page 31: ...minutes When you press ACCEPT the unit automatically checks the collected run time to see if it exceeds this limit If it does a PROGRAMMING ERROR message appears on the display for 5 seconds If this...

Page 32: ...rt with the small est locking collar Loosen the collar and push in the extension Tighten the smallest collar Continue with this procedure all the way down the mast un til the mast is collapsed 2 Unloc...

Page 33: ...F 6 2 0 1 0 8 k c a P y r e t a B r u o H o w T 2 0 0 0 1 0 8 c i t s e m o D E C r e g r a h C e l u d o M r e w o P 3 0 0 0 1 0 8 n a e p o r u E E C r e g r a h C e l u d o M r e w o P 6 3 0 1 0 4...

Page 34: ...2 Hour Battery Pack 6 5 lbs 2 9 kg 4 Hour Battery Pack 11 5 lbs 5 2 kg Power Module 2 lbs 0 9 kg Accessories Power Module Input 115 230 VAC 800 600 mA 47 63 Hz Output 15 Vdc 4 3 A Size 5 25 W x 7 5 D...

Page 35: ...anty repair orders will have a minimum fee whether the repair is authorized or not This fee in cludes handling administration and technical expenses for inspecting the instrument and providing an esti...

Page 36: ......

Page 37: ......

Page 38: ...16333 Bay Vista Dr Clearwater FL 33760 800 451 9444 727 530 3602 727 539 0550 FAX www sensidyne com Revision D Document No F PRO 3100...

Reviews: