![Sensica sensilight mini User Manual Download Page 24](http://html1.mh-extra.com/html/sensica/sensilight-mini/sensilight-mini_user-manual_1243504024.webp)
22
Guidance and manufacturer’s declaration – electromagnetic immunity
as per IEC 60601-1-2: 2007 -
Table 2
The SensiLight™Mini is intended for use in the electromagnetic environment specified below.
The customer or the user of the SensiLight™Mini should assure that it is used in such an
environment.
Immunity Test
IEC 60601
Test Level
Compliance level Electromagnetic
environment – guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 2 kV for power
supply lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to line(s)
*Not Applicable
*± 2 kV line(s) to
to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip
in U
T
) for 0,5 cycle
40 % U
T
(60 % dip
in U
T
) for 5 cycles
70 % U
T
(30 % dip in
U
T
) for 25 cycles
<5 % U
T
(>95 % dip
in U
T
) for 5 s
<5 % U
T
(>95 % dip in
U
T
) for 0,5 cycle
40 % U
T
(60 % dip in
U
T
) for 5 cycles
70 % U
T
(30 % dip in
U
T
) for 25 cycles
<5 % U
T
(>95 % dip in
U
T
) for 5 s
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
SensiLight™Mini requires continued
operation during power mains
interruptions, it is recommended that
the SensiLight™Mini be powered from
an uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for sensilight mini
Page 1: ...User Guide my permanent hair reduction Mini...
Page 2: ......
Page 27: ...25 Treatment date Treated area Comments Treatment log...
Page 28: ...26 Treatment date Treated area Comments Treatment log...
Page 29: ...27 Treatment date Treated area Comments Treatment log...
Page 30: ...28 Treatment date Treated area Comments Treatment log MD 04104 A0...