Maintenance 

 
The 

aSENSE RL

 is basically maintenance free. An internal self-adjusting calibration 

function takes care of normal long term drift associated to the CO

2

 sensor. To secure 

the highest accuracy, a time interval of five years is recommended between CO

2

 

calibrations, unless some special situations have occurred.  
 
A zero calibration can be performed by use of pure nitrogen or air that has passed 
through a chemical absorber and a PC together with the 

UIP software version 5 

(or 

higher) or by shortening of holes marked in figure 2 for at least 10 seconds. The 
sensor needs stable gas flow for at least 3 minutes before zero calibration. The Zero 
Calibration bag can be used to produce carbon dioxide free air. The software can be 
free downloaded from 

www.senseair.com

The RS232-cable and the zero calibration 

bag can be ordered from SenseAir. The cable is to be connected to the UART port 
slide connector (see Fig. 7). For change of control parameters and re-calibration of 
temperature this PC tool has to be used. Zero calibration can be done with PC and 

UIP software version 5

 or by shorting marked holes in figure 5. The check can be 

done on site without interfering with the ventilation system. 
 
A  Back-ground  calibration  is  an  easy  way  to  correct  a  sensor

’s  zero  point.  The 

sensor needs fresh air  (380-420 ppm CO

2

). Shorten holes marked  in  figure 5 for at 

least 10 seconds. 
 

This is for sensors with measuring ranges between 0-3000 ppm and 0-4%: 

When a zero calibration shall be executed a plastic tube with 2,2 mm outer diameter 
and 0,8 mm inner diameter shall be inserted in marked holes of the sensor. A plastic 
tubing is connected to the tube. The gas flow should  be between 0,3 and  0,8 l/min. 
The  sensor  needs  stable  gas  flow  for  at  least  3  minutes  before  zero  calibration. 
Shorten  marked  holes  for  at  least  10  seconds.  A  successful  calibration  is 
acknowledged by 5 flashes of the  yellow LED. If unsuccessful, please  release input 
and waits at least 10 seconds before repeating the procedure again. Make sure that 
the sensor environment is steady and calm! 
 

Gas inlet

 

Figure 5 Part of the PCB with holes for gas inlets marked.

 

Summary of Contents for aSENSE RL

Page 1: ...isplay The sensor is designed for wall mounting or mounting in ventilation ducts The unit is designed for connecting to Direct Digital Control DDC The linear output functions are pre programmed as CO2 and temperature transmitters with jumper selected outputs 0 10 V 2 10 V 0 20 mA 4 20 mA The measuring ranges can be modified from a PC Windows and use of the software UIP version 5 or higher together...

Page 2: ...ftware version 5 or higher Out 1 measuring signal of carbon dioxide concentration lowest value highest value Out 2 measuring signal of temperature lowest value highest value Out 3 ON OFF signal of demand of air quality The opening of the relay is controlled by the hysteresis of the relay See figure 4 Figure 2 The measuring signal of carbon dioxide Figure 3 The measuring signal of temperature Figur...

Page 3: ... for calculation 0 10 VDC 0 2000 ppm CO2 0 50 o C 32 122 o F CO2 value Volt 10 2000 Temperature value in o C Volt 10 50 Volt 5 Temperature value in o F Volt 9 32 2 10 VDC 0 2000 ppm CO2 0 50 o C 32 122 o F CO2 value Volt 2 8 2000 Temperature value in o C Volt 2 8 50 Temperature value in o F Volt 2 8 122 32 32 4 20 mA 0 2000 ppm CO2 0 50 o C 32 122 o F CO2 value mA 4 2000 16 Temperature value in o ...

Page 4: ...in position Current provides 0 4 20mA output range for Out 2 Connection in position Voltage provides 0 2 10VDC output range for Out 2 Table III Configuration jumpers for aSENSE RL FUNCTIONAL TEST aSENSE RL Self diagnostics The system contains complete self diagnostic procedures A full system test is executed automatically every time the power is turned on In addition constantly during operation th...

Page 5: ...ctions of temperature probe Check detailed status of measurements with UIP software version 5 and higher See Note 1 6 64 Memory Error Non fatal error during memory operations Check detailed settings and configuration with UIP software version 5 and higher 7 128 Warm Up state Is always set after power up or power failure Resets after restart sequence If it doesn t disappear in half a minute check p...

Page 6: ...rameters and re calibration of temperature this PC tool has to be used Zero calibration can be done with PC and UIP software version 5 or by shorting marked holes in figure 5 The check can be done on site without interfering with the ventilation system A Back ground calibration is an easy way to correct a sensor s zero point The sensor needs fresh air 380 420 ppm CO2 Shorten holes marked in figure...

Page 7: ...o damage the temperature probe in the sampling tube When putting the PCB back in the protective housing the probe must be gently positioned in the sampling tube Figure 6 The aSENSE RL for duct mounting The temperature probe is the black body in the sampling tube Figure 7 The aSENSE RL printed circuit board with CO2 sensors for measuring ranges 0 3000ppm and 0 4 ...

Page 8: ...by the CE marking Directive 93 68 EEC The product fulfils the following demand EN 61326 1 2006 Class B equipment Table 1 Basic immunity test requirements SenseAir AB headquarter Stationsgatan 12 Box 96 820 60 Delsbo SWEDEN Phone 46 0 653 71 77 70 Fax 46 0 653 71 77 89 E mail info senseair com Web site www senseair com ...

Reviews: