SenseAir aSENSE RL User Manual Download Page 4

 

 

 

 

 

Document 
UMA 149 
 
 

Rev 

Page 
4 (8)

 

 

Output Configurations

 

 
The sensors/controllers are supplied from the factory (unless otherwise ordered) with 
0...10VDC  linear  outputs  for  Out(1)  and  Out(2)  (see  Table  I).  If  other  options  are 
needed for the application, the output jumpers have to be configured before the unit 
is powered up. Each jumper selection is independent from the others, except for the 
“Start  point  selection”  jumper,  which  affects  both,  Out(1)  &  Out(2),  linear  outputs. 
Alternative  measuring  ranges  of  the  outputs  can  be  selected  with  the  software  UIP 
(version 5 or later).

 

 

 

Jumper 

 

Position 

 
Function 

 

 

Start point 

selection 

jumper 

 

0% 

 

 

20% 

 
Jumper top position provides 0Vdc or 0mA start point for Out(1), Out(2)  
(0-20mA or 0-10V). 
 
Jumper bottom position provides 2Vdc or 4mA start point for Out(1), Out(2) 
(4-20mA or 2-10V). 

 

 
 

Out(1) 

 

Current 

 

Voltage 

 
Connection in position “Current” provides 0/4-20mA output range for Out(1). 
 
Connection in position “Voltage” provides 0/2-10VDC output range for 
Out(1). 

 
 

Out(2) 

 

Current 

 

Voltage 

 
Connection in position “Current” provides 0/4-20mA output range for Out(2). 
 
Connection in position “Voltage” provides 0/2-10VDC output range for 
Out(2). 
 

Table III.  Configuration jumpers for aSENSE RL 

  

 
 

FUNCTIONAL TEST 

aSENSE RL

 

 

Self-diagnostics 

 
The  system  contains  complete  self-diagnostic  procedures.  A  full  system  test  is 
executed  automatically  every  time  the  power  is  turned  on.  In  addition,  constantly 
during operation, the sensor probes are checked against failure by checking the valid 
dynamic  measurement  ranges.  All  EEPROM  updates,  initiated  by  the  sensor  itself, 
as  well  as  by  external  connections,  are  checked  by  subsequent  memory  read  back 
and  data  comparisons.

 

These  different  system  checks  return  error  bytes  to  the 

system  RAM.  The  error  codes  are  available  by  connecting  a  PC  with  a  special 
RS232  cable  connected  to  the  UART  port  slide  connector.  The  error  codes  are 
shown  in  the  software  UIP  (version  5  or  later). 

Warm  up

  and 

Out of  Range

 are  the 

only bits that  are reset automatically after return to normal state.  All other error bits 
have to be reset manually after return to normal by power off/on. 
 
The  yellow  LED  flashes  if  an  error  has  been  detected.  If  a  fatal  error  has  been 
detected the yellow LED is lit. 
 

Summary of Contents for aSENSE RL

Page 1: ...isplay The sensor is designed for wall mounting or mounting in ventilation ducts The unit is designed for connecting to Direct Digital Control DDC The linear output functions are pre programmed as CO2 and temperature transmitters with jumper selected outputs 0 10 V 2 10 V 0 20 mA 4 20 mA The measuring ranges can be modified from a PC Windows and use of the software UIP version 5 or higher together...

Page 2: ...ftware version 5 or higher Out 1 measuring signal of carbon dioxide concentration lowest value highest value Out 2 measuring signal of temperature lowest value highest value Out 3 ON OFF signal of demand of air quality The opening of the relay is controlled by the hysteresis of the relay See figure 4 Figure 2 The measuring signal of carbon dioxide Figure 3 The measuring signal of temperature Figur...

Page 3: ... for calculation 0 10 VDC 0 2000 ppm CO2 0 50 o C 32 122 o F CO2 value Volt 10 2000 Temperature value in o C Volt 10 50 Volt 5 Temperature value in o F Volt 9 32 2 10 VDC 0 2000 ppm CO2 0 50 o C 32 122 o F CO2 value Volt 2 8 2000 Temperature value in o C Volt 2 8 50 Temperature value in o F Volt 2 8 122 32 32 4 20 mA 0 2000 ppm CO2 0 50 o C 32 122 o F CO2 value mA 4 2000 16 Temperature value in o ...

Page 4: ...in position Current provides 0 4 20mA output range for Out 2 Connection in position Voltage provides 0 2 10VDC output range for Out 2 Table III Configuration jumpers for aSENSE RL FUNCTIONAL TEST aSENSE RL Self diagnostics The system contains complete self diagnostic procedures A full system test is executed automatically every time the power is turned on In addition constantly during operation th...

Page 5: ...ctions of temperature probe Check detailed status of measurements with UIP software version 5 and higher See Note 1 6 64 Memory Error Non fatal error during memory operations Check detailed settings and configuration with UIP software version 5 and higher 7 128 Warm Up state Is always set after power up or power failure Resets after restart sequence If it doesn t disappear in half a minute check p...

Page 6: ...rameters and re calibration of temperature this PC tool has to be used Zero calibration can be done with PC and UIP software version 5 or by shorting marked holes in figure 5 The check can be done on site without interfering with the ventilation system A Back ground calibration is an easy way to correct a sensor s zero point The sensor needs fresh air 380 420 ppm CO2 Shorten holes marked in figure...

Page 7: ...o damage the temperature probe in the sampling tube When putting the PCB back in the protective housing the probe must be gently positioned in the sampling tube Figure 6 The aSENSE RL for duct mounting The temperature probe is the black body in the sampling tube Figure 7 The aSENSE RL printed circuit board with CO2 sensors for measuring ranges 0 3000ppm and 0 4 ...

Page 8: ...by the CE marking Directive 93 68 EEC The product fulfils the following demand EN 61326 1 2006 Class B equipment Table 1 Basic immunity test requirements SenseAir AB headquarter Stationsgatan 12 Box 96 820 60 Delsbo SWEDEN Phone 46 0 653 71 77 70 Fax 46 0 653 71 77 89 E mail info senseair com Web site www senseair com ...

Reviews: