Option 5a
Local alarm and
Nurse call
Local alarm off
Nurse call only
Option 5b
Figure 7
11
Local alarm on/off
If you are using the nurse call interface cord, you can set your TABS
unit for a local alarm and nurse call alarm
(Figure 7, Option 5a)
or
nurse call alarm only
(Figure 7, Option 5b)
. The nurse call interface
cord must be in place in the TABS monitor to make these selections.
If the cord is detached from the monitor, the unit will automatically
revert to the local alarm mode. If the monitor alarms for a period of
seven minutes, the local alarm will automatically turn on. To set the
nurse call only, move the number 3 rocker switch to the right position.
If the cord is unplugged from the call system and the alarm options
have been set to nurse call only, the TABS
®
monitor reverts to a local
alarm. The nurse call interface cord must be plugged into the nurse
call system at all times.
If the monitor alarms for a period of seven minutes, the unit will revert
to the local alarm mode. Before each use and periodically during the
day, check the cord to make sure it is securely in place. Failure to do
so may result in injury or death to a person in your care.
WARNING