User Manual
Z-SG2 / Z-SG2-L
ALL RIGHTS RESERVED. NO PART OF THIS PUBLICATION MAY
BE REPRODUCED WITHOUT PRIOR PERMISSION.
www.seneca.it
Doc: MI-005370-EN
Rev. 6
Page 23
CREATING A PROJECT
10.2.
ATTENTION!
It is necessary to set all the dip switches to OFF before sending the configuration to the device or the
configuration of the RS485 port will be consistent with the dip switch settings!
OPERATING MODE
10.2.1.
It allows to configure the basic operation of the device, can be set to factory calibration or to Calibration with
Sample weight:
10.2.1.1.
FACTORY CALIBRATION
It is used when a load cell with declared sensitivity (cell calibration) combined with the calibration of the
device's factory full scale is available.
In this mode the calibration consists only in acquiring the tare directly on the field with a direct measurement or
manually entering the tare value. For more information refer to chapters 11 and 14 of the manual.