- provide the signal corresponding to the
ex. 20 mA) and regulate the trimmer
full-scale (
Tr1
up to obtain the desidered indication.
Before using the instrument, verify in the rear label if the indicator has already been set by
the factory, for a particular input signal and a specific display view.
In this case a label is present on the calibration panel where the following informations are
listed:
-
input signal
-
the scale of display view.
If no label is present, the instrument has the following factory settings:
-Input: 4..20 mA
-View scale: 0..1000
Decimal Point Selection:
P1: one decimal, P2: two decimals
P3: three decimals
DECIMAL POINT SETTING
Three jumpers, placed in the rare of the frontal panel, define the position of the decimal
point:
FA TORY SETTINGS
C
P3 P2 P1