![SellEton Scales SL-7561 User Manual Download Page 12](http://html1.mh-extra.com/html/selleton-scales/sl-7561/sl-7561_user-manual_1237069012.webp)
10
F 2 Menu
Setting the Choosing a pad to calibrate, zero calibration, loading calibration, and saving
Step
Parameter
Display
Settings/Options
Enter F2 Menu
F 2
6
Pad Choice
SCALE 1
Choose a pad/scale to calibrate
Options: 1/2/3/4
7
Zero Calibration
CAL--0
0 = do not calibrate
1 = calibrate
Note: when 1 is selected the indicator will count down
from 9-0 and move on to the next setting
8
Calibration
SPAN-0
When ready for calibration, setting the span tells the
indicator the weight you will be calibrating the scale with
0 = do not set the span
1 = set the span
Use the arrow keys to set your calibration weight and
when done load your calibration weight to the scale
Once stable
press
to calibrate
The indicator will count down from 9-0 and move on to
the Saving parameter (Select 1 and press print to save
your calibration settings)
9
Save
SAVE-0
0 = do not save calibration
1 = save calibration
F 3 Menu
Step
Parameter
Display
Settings/Options
Enter F3 Menu
F
3
10
Axle Number
ALE-02
Set the number of axles (for use with accumulation
mode)
11
Cargo Number
SN--00
Set the cargo number
12
Printing Format
PF - 1
0 = No printing
1 = Normal printing format
2 = Accumulative printing format
* See page 13 for examples
13
Printing Method
P
M
- 1
0 = Manual
1 = Automatic
*Manual is recommended for the most precise results
14
Printing Coupon
PC - 1
Printing coupon numbers setting
Options: 1/2/3
15
Baud Rate
Baud-1
Options: 1/2/3/4/5/6/7/8
0 = 600 1 = 1200 2 = 2400
3 = 4800 4 = 9600 5 = 19200
6 = 38400 7 = 57600 8 = 115200
16
Communication
Settings
CP - 1
0 = Communication off
1 = Communication format 1 for PC
2 = Communication format 2 for second display
(YAO HUA MODEL)
3 = Communication format 3 for second display
(TOLEDO MODEL)