Bedienungsanleitung SDT43
BDA Serie SDT43 (Stand Jan. 2018)
Seite 44
LINEARIZATION HORIZONTAL TANK WITH A PARABOLIC END (CYLINDRICAL OR ELLIPTIC)
8.
Navigate to
Hor. Tank
. with the navigation button, and push to confirm.
9.
Two choices appear on the screen:
Input
and
Simulate
10.
Select
Input
, and push to confirm.
11.
Six choices appear on the screen:
Display
Drawing
Explanation
Length
L
The length of the tank
Height 1
H1
The height of the tank
Height 2
H2
The diameter of the tank
(with a cylindrical tank, this is equal to the height of the tank)
Height 3
H3
The height till the topside of the diaphragm (or weld-on nipple)
Height 4
H4
The length of
1
parabolic end of the cylinder
Fill Height
FH
The maximum percentage of filling of the tank
12.
Fill in each value, and confirm with the navigation button.
The entered value’s must be in
meters.
13.
Select
to save the setting.
14.
The transmitter will return to the main menu.
SIMULATION