INSTRUCTION MANUAL
CODE 80316
REV C
05/2020
Page: 11
J
.P. SELECTA s.a.u.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: [email protected] - website: http://www.jpselecta.es
4 Packing List
4.1
Standard package content
The standard equipment consists of the following components:
- Centrifuge MEDIFRIGER/MACROFRIGER BLT
7001477/7001486
- Centrifuge MEDITRONIC/MACROTRONIC BLT
7001476/7001485
- Network connection cable
7140
- Rotor nut wrench
5377
- Drainage nipple in refrigerated models
- Instructions manual
80316
3 Technical specifications
Voltage supply 115/230V 50/60 Hz as indicated in the
machine’s features plate.
Code:
7001476
7001477
7001485
7001486
Dimensions (mm): (Height/Width/Depth)
380x460x520
380x680x520
450x600x660
970x600x660
Power (W):
520
1230
720
1230
Max speed (rpm):
11000
15000
10000
12000
Max RCF.:
25470
(Depending on the rotor)
11066
15935
Weight (kg):
44
70
76
156
6 Installation
6.1
Location
• Place the equipment near a power socket suitable for consumption and over a flat
level surface and strong enough to support the equipment weight and possible
vibrations.
• The surface must allow the adherence of the centrifuge legs.
• Leave a free space of 30 cm. around it (see danger indication).
• This equipment can cause vibrations that could affect other nearby equipments.
• Do not install the centrifuge near heat sources or in places where it is exposed
to direct sunlight, as this may affect temperature regulation.
• Place the nipple provided in outlet (16) and connect to a hose to drain conden-
sates.
• Equipments with wheels, ensure that they are braked prior to start-up.
5 Introduction
The centrifuge has an induction motor maintenance free and a maximum noise level
of 60 dBA. It has an internal security camera and a bucket, stainless steel cover.
The electronic circuit is controlled by a microprocessor that controls the operating
parameters: speed, acceleration, braking, RCF, operating time and temperature on
models 7001486/7001477. It can store up to 10 operating programs.
FREE SPACE AROUND CENTRIFUGE
Safety zone:
Leave a free space of 30 cm. around
the centrifuge.
With the centrifuge running:
• Do not place hazardous substances in
the security area.
• The staff should be outside the safety
area.