MANUAL DE INSTRUCCIONES
CÓDIGO 80158 REV E
02/2021
Pag.: 32
J
.P. SELECTA s.a.u.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 93 770 08 77 Fax 34 93 770 23 62
e-mail: [email protected] - website: http://www.grupo-selecta.com
11 Digestion on C.O.D. determination
• Add 5.0ml of potasic dichromat and some glass balloons to regulate ebullition
to the sample and it is homogenized carefully.
• Add slowly 15ml. of silver sulphate - sulphuric acid, shaking the tube carefully
with a circular movement, cooling down with cool water of the tap or an ice bath,
in order to avoid any loss of organically volatile substances.
• Place the tubes dully prepared into the rack tubes holder on their support.
With the help of a rack, insert it into the heating block.
• Join on each tube a refrigerant in order to fit correctly the polished gaskets.
• Program the operating temperature (170-200ºC) and the length time (around
2 hours) of the back flowing into the temperature regulator and RAT time.
• Press on «START» and at this moment the metallic block will heat where
the tubes are placed. This will be done in a progressive way and when the
preselectionated temperature is reached, the time of back flowing will be counted.
• Once the time is finished, the heating in the metallic block stops automatically
and an acoustic sound is heard. At this moment, you should press on «STOP»
in order to finish the process.
• Let the tubes cool down for a while and clean the projections that have been
produced in the inner wall of the refrigerants into the sample recipient with a
little volume of distilled water.
• Separate the refrigerants from the tubes, and the tubes are extracted with the
help of the rack, that is placed on the support. The sample is diluted with 75 ml.
of distilled water and it is cooled down at ambient temperature.
Notice to customers:
The product is made up of various components and various materials that must
be recycled or, failing that, deposited in the corresponding debris removal sites
when the product’s life has been completed or when otherwise it is necessary
to dispose of it. To do this, the end user who acquires the product must know
the current regulations of each municipality and / or locality based on the waste
electrical and electronic equipment. The user who acquires this product must
be aware of and responsible for the potential effects of the components on the
environment and human health as a result of the presence of hazardous sub-
stances. Never place the product in a conventional container of citizen scope if
a previous dismantling and knowledge of the components that incorporates. If
you do not know the procedure to follow, consult with the city council for more
information.