18
Setting the delay time for the lid open alarm
The time before the audible warning signal sounds after the lid
has been opened can be adjusted.
Press
for 5 seconds. Display =
Display =
Display =
Display =
Display =
Display = (minute)
Setting range = 1 - 5 minutes
Use buttons
and
to select the desired setting.
Display =
Press
for 5 seconds.
The electronic control system will switch back to normal operat-
ing mode.
Audible warning signal settings
The audible warning signal will be muted for the current alarm after
the button
has been pressed. Complete the following steps if
you want the audible warning signal to reactivate automatically.
Press
for 5 seconds. Display =
Display =
Display =
Display =
Display =
Display =
Display =
Display =
Display =
Automatic reactivation of the audible warning signal is now active.
The time before the audible warning signal sounds again must be set.
Display =
Display = Time in minutes after which the audible
warning signal will sound again after the
button
has been pressed. Setting
range = 1 - 120 minutes.
Use buttons
and
to select the desired setting.
Display =
Press
for 5 seconds.
The electronic control system will switch back to normal operat-
ing mode.
Deactivating the audible
warning signal function
The audible warning signal function can be completely deacti-
vated if necessary.
Note
In this case, the sentence stated in these operating instructions
"
The audible warning signal sounds
" must be skipped when
reading the section in question.
Press
for 5 seconds. Display =
Display =
Display =
Display =
Use buttons
and
to select the desired setting.
= audible warning signal function activated
= audible warning signal function deactivated
Display =
Press
for 5 seconds.
The electronic control system will switch back to normal operat-
ing mode.