INSTRUCTIONS MANUAL CODE 80021 REV F
May/2012
(It can be modified without notice) Page.: 14
J
.P. SELECTA s.a.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 93 770 08 77 Fax 34 93 770 23 62
e-mail: [email protected] - website: www.jpselecta.es
COMMENTS:
* The thermostat scale is intended as a guide, the real temperature
level inside the incubator is the one that the thermometer (12) shows.
* Do not use the incubator at a lower temperature than ambient +5ºC,
because the thermostat will not operate correctly.
* All the temperature regulations must always be made in ascent.
Before choosing any value, the knob must be turned left to the end of
the scale. Then a more exact regulation is obtained.
* Place the load in the incubator trying to leave spaces for the air
circulation. Do not load the incubator more than 70% of its useful
chamber volume.
ADJUSTING THE SAFETY THERMOSTAT:
To adjust the safety thermostat temperature do as follows:
1. With the incubator connected to the mains, turn the safety thermostat
control knob (4) clockwise to the end of the scale.
2. Select the desired safety temperature with the temperature control
knob (10). This temperature has to be minimum at 3ºC above the
working temperature.
3. With the door closed, turn on the incubator with the main switch (1).
4. When the incubator is stabilized at the above selected temperature,
turn the safety thermostat control knob (4) slowly to the left until it
clicks, the red indicator lamp (6) lights up.
5. Stop the incubator and leave it to cool with the door open during 15
minutes.
6. Block the temperature control knob (10) with the blockage lever (9)
to avoid accidentally moving the knob.
7. Select the working temperature with the thermostat control knob
(10) (minimum at 3ºC below the desired safety temperature).
8. Start the incubator once again switching it on (1).
9. Block the thermostat knobs with their corresponding blockage devices
(5 and 9).
10. Close the door and turn the incubator on again with the main switch
(1). Once the temperature is stabilized, the incubator is ready to
work.
NOTE: The safety thermostat is useful when the user works habitually
at the same temperature.