EM00136061
rev. 2.5
EN
4
Display during Operation
Display during Setting (MODE key)
mA
F
F
F
4-20
Lev
Stop
P100%
Operating mode
mA
Alarms and statuses
Lev = Level alarm
Flw = Flow alarm
FLOW sensor
status
Programmed value
Programmed interval value
(default: 4-20)
Pump status
Empty = pump in start
Stop = pump stationary
Paus = pump in pause
Current dosage value
(depends on the selected unit
of measurement)
Percentage, Frequency,
l/h, Gph, ml/m
Low
0
0.0 mA
P100%
Operating mode
Displays the Low and High
value in sequence
Input mA
value
Current dosage value
Makes it possible to modify the
maximum dosage value by pressing
the + or – buttons at the same time
Paragraph 3 - Dosage Proportional to Signal 20-4/0 mA
Programming
Operation
+
-
-
-
+
-
+
-
Pump Funtions
Configuration
Pump Functions
20-4/0 mA
Low
20.0mA
High
4.0mA
PROG
20-4/0 mA
20-4/0 mA
The pump proportionally doses at a signal of 20-4(0) mA. On the basis of
the factory settings, the pump interrupts dosage at 20mA and doses at the
maximum set frequency when it receives 4 mA. These two values can be
modified during programming. The maximum frequency can be modified
during operation, by pressing the
keys at the same time to
increase the flow, or the
keys to decrease it.
Display during Operation
Display during Setting (MODE key)
mA
F
F
F
20-4
Lev
Stop
P100%
Operating mode
mA
Alarms and statuses
Lev = Level alarm
Flw = Flow alarm
FLOW sensor
status
Programmed value
Programmed interval value
(default: 20-4)
Pump status
Empty = pump in start
Stop = pump stationary
Paus = pump in pause
Current dosage value
(depends on the selected unit
of measurement)
Percentage, Frequency,
l/h, Gph, ml/m
High
0
0.0 mA
P100%
Operating mode
Displays the Low and High
value in sequence
Input mA
value
Current dosage value
Makes it possible to modify the
maximum dosage value by pressing
the + or – buttons at the same time
Paragraph 4 – Proportional to External Impulses (multiplication)
Programming
Operation
+
-
-
-
-
+
-
+
-
+
-
Pump Funtions
Configuration
Pump Functions
Multiply (1:n)
n
4
Timeout
0s
Memory
Off
PROG
Multiply (1:n)
Multiply (1:n)
Multiply (1:n)
The pump doses in proportion to an external signal (i.e.: impulse launch
counter). With every signal received, the pump makes the programmed
“n” number of strikes. The pump automatically sets the dosage frequency,
adapting it to the time that passes between two successive signals. It is
possible to set the time (timeout) in seconds, beyond which the pump
resets the interval count in order to avoid dosages over excessively long
times. The pump has a memory function, which signals the reception of a
signal during dosage. If set to Off, it merely sends a signal, if set to On it
sends a signal and memorises the impulses, then executes them when it
has finished receiving signals.
The value of “n” can be modified during operation, by pressing the
keys at the same time to increase the flow, or the
keys to
decrease it.