Performance Characteristics
43
Precision:
To maximize your chances of an accurate comparison between meter
and laboratory results, follow a few basic guidelines:
Before going to the lab
·
Perform a control solution test to make sure the meter is working
properly.
·
Do Not eat for at least eight hours before you test your blood.
Take your meter with you to the lab.
·
While at the lab
Conduct your meter test within 15 minutes of the lab test.
·
Use fresh capillary blood obtained from the fingertip, venous blood,
·
neonatal blood.
Follow all instructions in this Owners Booklet for performing a
·
Control Solution
av
2.20mmol
/
L
CV
=
3.69%
Intermediate
Precision
Control Solution
av
4.89mmol
/
L
CV
=
2.82%
Control Solution
av
7.19mmol
/
L
CV
=
1.97%
Blood
av
2.55mmol
/
L
CV
=
4.3%
Blood
av
4.79mmol
/
L
CV
=
4.2%
Blood
av
7.34mmol
/
L
CV
=
2.3%
Blood
av
11.42mmol
/
L
CV
=
2.6%
Blood
av
20.5mmol
/
L
CV
=
1.6%
Repeatability
Control Solution
av
10.99mmol
/
L
CV
=
1.77%
Control Solution
av
18.09mmol
/
L
CV
=
1.36%
Table3. System accuracy results for glucose between 0.5 mmol/L (9 mg/dL)
and 33.3 mmol/L (600 mg/dL)
Within
(Within ±15mg/dL or ±15%)
±0.83mmol/L or ±15%
596/600(99.3%)
This study shows that the system compares well with a laboratory
the minimum acceptable performance criteria
method and meets
defined in EN ISO 15197 : 2015.
blood glucose test with your meter.