15
1
Switch off the instrument and disconnect the power cable from the instrument connector.
4
Remove both tubes from the elbow fittings.
2
Locate the rotor of the peristaltic pump.
3
Pull out the rotor of the peristaltic pump, holding
inward the two plastic fins indicated by the arrows
and at the same time, extract the rotor.
5
Pulled out the old rotor, replace it with a new one
doing reverse operations up to here described,
paying attention to the correct insertion of the
tubes that must be crossed.
7.2 PERISTALTIC PUMP ROTOR REPLACEMENT
For the replacement of the peristaltic pump rotor, proceed as follows