• DO NOT USE the test leads above maximum
ratings of CAT. environment, voltage and current,
that are indicated on the probe and the probe tip
guard cap.
• DO NOT USE the test leads without the probe tip
guard cap in CAT III and CAT IV environments.
• Probe assemblies to be used for MAINS meas-
urements shall be RATED as appropriate for
MEAS UREMENT CATEGORY III or IV accord-
ing to IEC 61010-031 and shall have a voltage
RATING of at least the voltage of the circuit to be
measured.
• Disconnect circuit power and discharge all high
voltage capacitors before testing resistance,
continuity, diodes, or capacitance.
• Do not use a flexible current sensor if the inner
copper wire of the insulation of the flexible cord
is visible.
• De-energize the installation under test or wear
suitable protective clothing during fitting and
removal of the Flexible Current Probe.
• Do not apply around or remove from UNINSU-
LATED HAZARDOUS LIVE conductors, which
may render electric shock, electric burn, or arc
flash.
Caution
CAUTION identifies conditions and actions that
could damage the Meter, the equipment under
test, or cause permanent loss of data.
• Remove the batteries if the Meter is not used for
an extended period of time, or if stored in
temperatures above 50 °C. If the batteries are not
removed, battery leakage can damage the Meter.
• Do not expose Meter to extremes in temperature
of high humidity.