![SEFRAM SP260B User Manual Download Page 20](http://html1.mh-extra.com/html/sefram/sp260b/sp260b_user-manual_1233381020.webp)
9
1-3 Electrical Specifications
Accuracy is given as ± (% of r counts of least
significant digit) at 23°C ± 5°C, with relative humidity
less than 80% R.H. Accuracy is specified for a period
of one year after calibration.
Position Error of Clamp:
Add ±1.5% of LCD reading.
Ampere
Range Output
Ratio
Accuracy
100A
ġ
10mV/A
ġ
±
ĩ
2.0% +0.4A)
600A
ġ
1mV/A
ġ
±
ĩ
2.0% +2A)
Frequency Response :
DC,AC 40Hz to 400Hz
Output Impedance :
< 10k
Overload Protection :
AC/DC 600A
VoltSeek
Voltage Range of Sensitivity :
90V to 1000V
(At the top edge of the meter)
Frequency Response :
50 to 60Hz (Sine Wave)
Summary of Contents for SP260B
Page 11: ......