30
M35X6BF-M-00
(2) Current
* 3526BF : 599.9A ; 3536BF : 999.9A
** The measured value <1000dgt, add 5 dgt to the accuracy.
Position Error :
±1% of reading.
AC Conversion Type and additional accuracy is same as AC
Voltage.
AC+DC Arms Accuracy :
Same as ACA spec + DCA spec.
-
For better measurement accuracy of high current and the
constraint of temperature increasing of maximum range
600A/1000A AC, do not measurement more than 10min. and
have rest time with 30min. at least in between every measure-
ment (for 3516BF).
-
DCA affected by the temperature and the residual magnetism.
Press HOLD key > 2sec to compensate it.
3526BF/3536BF
Function
Range
Accuracy
DCA
99.99A
± (1.5% + 0.2A)
599.9A/999.9A*
± (1.5% + 5dgt)
**
ACA
0.10A ~ 99.99A
50 ~ 60Hz ± (1.5% + 5dgt)
**
>60 ~ 400Hz ± (2% + 5dgt)
**
599.9A/999.9A*
HFR
ACA
0.10A ~ 99.99A
50 ~ 60Hz ± (1.5% + 5dgt) **
>60 ~ 400Hz ± (5% + 5dgt)
**
599.9A/999.9A*
* 3516BF : 599.9A
** The measured value <1000dgt, add 5 dgt to the accuracy.
3516BF
Function
Range
Accuracy
ACA
99.99A
50 ~ 60Hz ± (1.5% + 5dgt)
**
>60 ~ 400Hz ± (2% + 5dgt)
**
599.9A/999.9A*
HFR
ACA
0.10A ~ 99.99A
50 ~ 60Hz ± (1.5% + 5dgt) **
>60 ~ 400Hz ± (5% + 5dgt)
**
599.9A/999.9A*