17
M35X6BF-M-00
Phase Rotation
Line 2
Line 3
Line 1
NOTE :
- Connect the supposed three phase of power source as shown
above.
- The test is only available while the system frequency is stable.
Step 1. Set the rotary switch to the "W" position.
Step 2. Using the MODE button to choose the "RST" mode
Step 3. Connect the Red test lead to the supposed phase
Line 1, and the Black test lead to the supposed phase
Line 3.
a. if volt > 1050V, it will display “OLU” and flash; if volt <30V,
it will display “LoU”
b. If the frequency > 65Hz or < 45Hz, it will display “outF”
and flash.
c. If it is normal, then it will display “L1” and flash for about
3 sec.