EcoStar
Settings
9
❶
This screen appears:
Press
simultaneously
the hidden key
and
for one second
Level of prescribed pressure
Increase or decrease the value displayed
using the
button or the
button.
❷
Press
Ramp time
Increase or decrease the value displayed
using the
button or the
button.
❸
Press
Comfort Pressure
This parameter is only adjustable if the
ramp time is greater than 0.
Increase or decrease the value displayed
using the
button or the
button.
For further information, refer to "Ramp
feature” on page 7.
❹
Press
Hour counter value (device operating
time).
The value displayed on the right below
represents the minutes. The hour meter
value cannot be changed.
❺
Press
Value of the compliance counter (time
during which the patient actually breathed
with the mask). The value displayed on
the right below represents the minutes.
This counter can only be reset to zero. To
do this, press the hidden key, then keeping
it held down, press the
button or the
button.