Electrical Lock
Mechanical Lock
General Purpose Rad System
Operation
OM-0273R2
12
3.4
WALL BUCKY STAND / WALL SUPPORT STAND
The Wall Bucky Stand / Wall Support Stand enables radiographic operations
at different positions within the range of the Vertical Carriage travel.
ON / OFF:
The Wall Bucky Stand / Wall Support Stand is turned ON / OFF when
the X-ray Generator is turned ON / OFF. This only applies equipment with
Electrical Lock.
VERTICAL CARRIAGE ASSEMBLY:
The carriage slides vertically along the
Column.
In the Unit with Electrical Lock, the lock is released by pressing the Carriage
Handle Button in order to allow vertical movement and positioning with respect
to patient. Release button to lock in position. As well, the vertical movement is
locked when the equipment is turned OFF.
In the Unit with Mechanical Lock, the lock is released by turning the Handle to
allow vertical movement and positioning with respect to patient. Tighten the
Handle to lock in position. Handle can be installed at any side of the carriage.
CASSETTE TRAY :
Pull out Cassette Tray to load Cassette Film according to
cassette size and orientation. Place manual clamps at the corresponding
numbered notch, open the automatic clamp and insert Cassette Film. This
Cassette Tray accepts all standard Cassette Film sizes.
Opening of the Cassette Tray can be located at right or left side
of the Column as per Unit model.
BUCKY FRONT PANEL:
The Front Panel is made of a special thermostable
resin with a very low absorption level. It includes lateral aluminium profiles with
grooves for standard accessories. Also provides a Chin Rest support at the
upper side. The panel shows AEC detector Areas and Cassette Film sizes.
Note
.