General Purpose Rad System
Operation
OM-0273R2
9
SECTION 3
OPERATION
USE THE EQUIPMENT HAND-GRIPS TO CONTROL AND
DRIVE THE UNIT MOVEMENTS, NEVER PUSH DIRECTLY ON
THE BUCKY, X-RAY TUBE OR COLLIMATOR.
MONITOR THE SYSTEM MOVEMENTS WITH SPECIAL
CARE. AVOID ANY IMPACT OF THE SYSTEM ON FLOOR,
CEILING OR OTHER ELEMENTS IN THE ROOM. IT MAY
CAUSE SERIOUS DAMAGE TO THE EQUIPMENT.
MONITOR WITH SPECIAL CARE THE PATIENT POSITION
(HANDS, FEET, FINGERS, ETC.) AND USE THE PATIENT
HAND-GRIPS TO AVOID INJURY TO PATIENT CAUSED BY
TABLE-TOP MOVEMENTS. PATIENT HANDS MUST BE KEPT
AWAY FROM MOBILE COMPONENTS OF THE UNIT.
OPERATOR SHOULD NEVER MOVE TABLE-TOP FROM
LATERAL SIDE (PATIENT HEAD OR FEET). IT MAY CAUSE
FINGERPINCH UNDER THE TABLE-TOP.
INTRAVENOUS TUBING, CATHETERS AND OTHER PATIENT
CONNECTED LINES SHOULD BE ROUTED AWAY FROM
MOVING EQUIPMENT.
3.1
POWER ON / OFF FOR THE X-RAY ROOM
All the equipments of the X-Ray Room should be powered by the same Room
Electrical Cabinet where the X-ray Generator is connected. When the Control
Console of the X-ray Generator is Turned ON / OFF, the whole System will be
turned ON / OFF.
IN THE EVENT OF AN EMERGENCY, PRESS FORCIBLY THE
“EMERGENCY OFF SWITCH” (RED MUSHROOM-SHAPED
SWITCH) OF THE ROOM ELECTRICAL CABINET.