HME-PREMIO30-ENG ENGLISH 11/28
H
OW TO USE THE
P
REMIO
30
LASER DUO
Use of commands or settings or carrying out procedures
other than those described in this manual may result in
exposure to dangerous irradiation.
Before starting
a.
To recharge the
Premio 30 laser duo:
The
Premio 30 laser duo
cannot be used during charging.
•
Remove the laser key,
•
Insert the charger plug into the instrument,
•
Connect the charger to the power supply.
Power indicator: (extract from the documentation *)
• Yellow: Charging
(steady then flashing at the end of charging)
• Solid Green: Fully charged the battery.
• Flashing green:
The PREMIO is not connected.
See charger manual * for:
•
LED off or blinking red: Battery / charger fault
• Solid Flashing red: Temperature fault
*: See the attached manufacturer's manual or on
https://www.mascot.no/downloads/user-manuals/battery-
chargers/ “chargers_li
-
ion”)
In the event of a
Battery fault
: See chapter Troubleshooting.
b.
To check correct operation of the
Premio 30 laser duo:
-
Remove the laser key,
-
Introduce the test laser plug into the
instrument,
-
Connect the laser key to the test laser,
-
Press on the
key.