11.7 EMR system connection via seca connect 103 software
For more information on using the
seca connect 103
software, see the
seca 103/452
system instructions for use.
Fault
Cause
Remedy
Data transmission cannot
be set up between device
and
seca connect 103
Software versions not compatible
Administrator:
Use a compatible version of the
seca connect 103
BIA measuring mat activated.
EMR system connections are possible
only with the BIA measuring mat
deactivated
Deactivate measuring mat, more
information is available here:
Unable to find user ID
No connection to the server
Administrator:
Check server connection
User ID not recognized by the server
Ensure that user ID is valid
Unable to find patient ID
No connection to the server
Administrator:
Check server connection
Patient ID not recognized by the server
Ensure that patient ID is valid
Unable to scan IDs
Scanner not connected correctly
Ensure that the USB stick is correctly
located in the USB interface
Barcode scanner incompatible
Use compatible barcode scanner
Scanner defective
Replace scanner
There is no request for a
password following input
of the EMR user ID. The
Vital signs tab is displayed
immediately
Not a malfunction:
The device connection to the EMR
system has been configured so that no
password needs to be entered.
Administrator:
Modify configuration if necessary
Unable to find EMR system
patient data
No patient file available in the EMR
system
Create patient file directly in the EMR
system
Error scanning patient ID
Scan patient ID again
No network connection to the EMR
system
Administrator:
Check the network connection and set up
again if necessary
Login using EMR user ID
not possible
EMR system connection interrupted,
device set up so that no measurements
are possible
Administrator:
If possible, allow measurements with the
EMR system connection interrupted
Blood pressure values do
not appear in the EMR
system
Multiple measurement was performed
For current measurement:
•
and
Administrator:
• Specify Individual measurement as preset
Not possible to submit
measuring results to the
EMR system
Network cable not plugged in correctly
Check whether the network cable is
connected correctly and that the
connectors are firm in their sockets
Network cable defective
Replace network cable
Device outside WiFi range
Bring device into WiFi range
No network connection to the EMR
system
Administrator:
Check the network connection and set up
again if necessary
Network connection interrupted