8. HYGIENE TREATMENT
WARNING!
Electric shock
The device is not de-energized when the on/off key is pressed and the
display goes out. Use of fluids on the device may cause electric
shock.
►
Ensure that the device is switched off before performing any hygiene
treatment.
►
Disconnect the power supply connector before performing any
hygiene treatment.
►
Before each hygiene treatment, take the rechargeable battery out of
the device (if present and removable).
►
Ensure that no fluids penetrate the device.
NOTICE!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
►
Use only disinfectants free of chlorine and alcohol which are explic
-
itly suitable for acrylic sheet and other sensitive surfaces (active in
-
gredient: quaternary ammonium compounds, for example).
►
Do not use caustic or abrasive detergents.
►
Do not use organic solvents (e.g. white spirit or petroleum spirit).
►
Use disinfectants including the active ingredient 70 % isopropanol
only for measuring accessories for measuring vital signs.
8.1 Cleaning
►
Clean the device and its accessories as described in the table:
Removing/fitting probe holder (devices
Removing/fitting the magazine holder
(devices with in-ear thermometer)
Component (depending on version)
Interval
Cleaning
seca mBCA 525
: Monitor with storage
compartment
seca mVSA 535
: Monitor with
SmartBucket
As required
1. Remove all the measuring accessories (measuring
devices and consumables) from the device (depending
on version)
•
•
2. Moisten a soft cloth with a soap solution
3. Wipe over all surfaces
4. Allow to air-dry for approx. 30 minutes
Measuring mat and electrode cables
As required
1. Moisten a soft cloth with a mild soap solution
2. Clean measuring mat and electrode cables
3. Allow to air-dry for approx. 30 minutes
Blood pressure cuff and compressed air
tube
As required
1. Moisten a soft cloth with a mild soap solution
2. Clean blood pressure cuff and compressed air tube
3. Rinse thoroughly with water
4. Allow to dry at room temperature