![Seca mBCA 525 Instructions For Use Manual Download Page 104](http://html1.mh-extra.com/html/seca/mbca-525/mbca-525_instructions-for-use-manual_1230118104.webp)
104 •
seca mVSA 535
/
seca mBCA 525
1
7-
10
-0
5-
35
3-
00
2g
_1
0-
20
20
B
8.2 Disinfecting
1. Follow the instructions for use for the disinfectant.
2. Disinfect the device and its accessories as described in the table:
8.3 Sterilizing
The device and accessories must not be sterilized.
Component (depending on equipment)
Interval
Disinfecting
seca mBCA 525
: Monitor with storage
compartment
seca mVSA 535
: Monitor with
SmartBucket
As required
1. Remove all the measuring accessories (measuring
devices and consumables) from the device (depending
on version)
2. Moisten a soft cloth with disinfectant (active ingredient:
quaternary ammonium compounds)
3. Wipe over all surfaces
4. Allow to air-dry for approx. 30 minutes
Measuring mat and electrode cables
Before
and
after
a
measurem
ent
1. Moisten a soft cloth with disinfectant (active ingredient:
quaternary ammonium compounds)
2. Wipe over measuring mat and electrode cables
3. Allow to air-dry for approx. 30 minutes
Blood pressure cuff and compressed air
tube
As required
1. Moisten a soft cloth with disinfectant (active ingredient:
70 % isopropanol)
2. Wipe over cuff and compressed air tube
3. Rinse thoroughly with water
4. Allow to dry at room temperature
Temperature probes (red/blue) with cable
As required
1. Discard probe cover and dispose of it
2. Moisten a soft cloth with disinfectant (active ingredient:
70 % isopropanol)
3. Wipe over temperature probe
4. Shake out temperature probe so that no liquid remains
in it
5. Allow to air-dry for approx. 30 minutes
Probe holder (red/blue)
As required
1.
2. Moisten a cotton bud with disinfectant (active
ingredient: 70 % isopropanol)
3. Wipe over the inner surfaces of the probe holder
4. Allow to air-dry for approx. 30 minutes
In-ear thermometer, cable, measuring
head and lens
As required
1. Discard probe cover and dispose of it
2. Wipe over in-ear thermometer several times with a
cloth moistened with alcohol (70 % isopropanol) so that
all surfaces are visibly moist for at least one minute
3. Make sure that the lens at the tip of the measuring
head does not have any fingerprints or marks on it
4. Allow the in-ear thermometer to dry completely in air
Magazine holder for probe covers (in-ear
thermometer)
As required
2. Moisten a cotton bud with disinfectant (active
ingredient: quaternary ammonium compounds)
3. Wipe over surfaces
4. Allow to air-dry for approx. 30 minutes
SpO
2
sensor with cable
Follow the manufacturer’s instructions for use
Patient cable for SpO
2
sensor
Follow the manufacturer’s instructions for use